supplemental oxygen

supplemental oxygen

EN

tubing to the open end of the adapter and the other end of the air supply tubing to the mask.

2.If your mask is equipped with an oxygen port, your physician may request that you apply oxygen directly to your mask.

WARNING-Oxygen use

Oxygen supports combustion. To avoid possible physical injury, do not smoke while using this device with supplemental oxygen. Do not use this device near hot objects, volatile substances or sources of open flames.

Always turn on the IntelliPAP before turning on the oxygen source. Turn the oxygen source off before turning off the IntelliPAP.

This device is not suitable for use in the presence of a flammable anesthetic mixture with air or oxygen, or nitrous oxide.

At a fixed flow rate of supplemental oxygen, the inhaled oxygen concentration will vary depending on the pressure setting, patient breathing pattern, mask selection, and leak rate. This warning applies to most types of CPAP devices.

Never allow the oxygen source to run continuously while connected to the IntelliPAP if the device is not in use. If the IntelliPAP is not being used, turn off the oxygen flow.

CAUTION–Oxygen is a prescription gas and should only be administered under the supervision of a physician.

CAUTION–The flow setting of the oxygen source must be specified by a physician.