DeVillbiss Air Power Company manual A-DV54, troubleshooting, Problem, Probable Cause, Remedy

Models: DV54

1 76
Download 76 pages 46.99 Kb
Page 22
Image 22
22A-DV54

EN

troubleshooting

Problem

Probable Cause

Remedy

 

 

 

The display

An error has occurred in the

Contact your home care provider

shows a device

device and requires servicing.

for service.

fault.

 

 

 

 

Symptoms of

1.

Air filter may be occluded.

1.

Clean or replace the air filter

OSA have

 

 

 

and relocate device away from

recurred.

 

 

 

drapes and other dusty surfaces.

 

2.

Sleep apnea condition has

2.

Contact your physician.

 

 

changed.

 

 

Your skin

1.

Headgear is too tight or

1.

Loosen the headgear to reduce

becomes

 

improperly adjusted.

 

the contact pressure on your

irritated where

 

 

 

face.

the mask

2.

Your mask may not be

2.

Contact your physician or home

touches your

 

properly sized or is not the

 

care provider.

face.

 

most appropriate shape for

 

 

 

 

you.

 

 

Dryness of

Inadequate humidity.

1.

Add a DeVilbiss DV5 Heated

throat or nose.

 

 

 

Humidifier.

 

 

 

2.

Increase the heater setting on

 

 

 

 

the humidifier.

Condensation of

1.

Your humidification setting is

Reduce the humidifier heater

water is

 

too high.

setting and/or increase the

collecting in the

2.

The room temperature

temperature of the room.

hose causing a

 

fluctuates from higher to

 

 

gurgling noise.

 

lower levels through the

 

 

 

 

night.

 

 

Air from flow

1.

The air filter(s) is/are dirty.

1.

Refer to the Cleaning section for

generator seems

 

 

 

instructions on filter

to be too warm.

 

 

 

maintenance.

 

2.

The air inlet port is blocked.

2.

Unblock the air inlet.

 

3.

The room temperature is too

3.

Reduce the room temperature.

 

 

high.

 

 

 

4.

The IntelliPAP is located near

4.

Locate the device away from

 

 

a heater.

 

the heat source so the air

 

 

 

 

being drawn into the device is

 

 

 

 

at room temperature.

 

 

 

If the problem persists contact

 

 

 

your home care provider.

Nasal, sinus or

You may have a reaction to air

Discontinue use and contact your

ear pain,

flow pressure.

physician.

possibly with

 

 

 

 

runny nose.

 

 

 

 

22A-DV54

Page 22
Image 22
DeVillbiss Air Power Company manual A-DV54, troubleshooting, Problem, Probable Cause, Remedy