30 Chapter6Troubleshooting

Error Code Details

**34 • Thesensormaynothaveworkedcorrectly.
• Turnoffthemainpower,waitforafewminutesbeforeturningthepowerback
on,andthenexecuteSelfCalibration/GrayscaleChec kagain.
**94 • <ScreenSize>is“Enlarged”or“Full”,andSelfCalibrati onorGrayscaleCheck
cannotbeexecuted.
• Set<ScreenSize>to“Normal”.
**95 • TheexecutionconditionsforSelfCalibrationorG rayscaleCheckmaynotbe
appropriate.
• Checktheexecutionconditionsandresetthemifnece ssary.