44 Appendix
EMC Information

Essential performance of RadiForce series is to display images and operate functions normally.

CAUTION

The RadiForce series requires special precautions regarding EMC and need to be installed, put into

service and used according to the following information.

Do not use any cables other than the cables that provided or speci ed by us.

Using other cables may cause the increase of emission or decrease of immunity.

Do not put any portable and mobile RF communications equipment close to the RadiForce series.

Doing so may affect the RadiForce series.

The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent or

stacked use is necessary, the equipment or system should be observed to verify normal operation

in the con guration in which it will be used.

Anyone who connects additional equipment to the signal input part or signal output parts,

con guring a medical system, is responsible that the system complies with the requirements of IEC/

EN60601-1-2.

Guidance and manufacturer’s declaration - electromagnetic emissions

TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.

Emission test Compliance Electromagnetic environment - guidance

RFemissions
CISPR11/EN55011
Group1 TheRadiForceseriesusesRFenergyonlyforitsinternalfunction.
Therefore,itsRFemissionareverylowandarenotlikelytocauseany
interferenceinnearbyelectronicequipment.
RFemissions
CISPR11/EN55011
ClassB TheRadiForceseriesissuitableforuseinallestablishments,includingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Harmonicemissions
IEC/EN61000-3-2
ClassD
Voltageuctuations/
ickeremissions
IEC/EN61000-3-3
Complies

Guidance and manufacturer’s declaration - electromagnetic immunity

TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.

Immunity test IEC/EN60601 test

level

Compliance level Electromagnetic environment - guidance

Electrostatic
discharge(ESD)
IEC/EN61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Electricalfast
transient/burst
IEC/EN61000-4-4
±2kV
forpowersupplylines
±1kV
forinput/outputlines
±2kV
forpowersupplylines
±1kV
forinput/outputlines
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Surge
IEC/EN61000-4-5
±1kVline(s)toline(s)
±2kVline(s)toearth
±1kVline(s)toline(s)
±2kVline(s)toearth
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC/EN61000-4-11
<5%UT(>95%dipin
UT)for0.5cycle
40%UT(60%dipinUT)
for5cycles
70%UT(30%dipinUT)
for25cycles
<5%UT(>95%dipin
UT)for5sec
<5%UT(>95%dipin
UT)for0.5cycle
40%UT(60%dipin
UT)for5cycles
70%UT(30%dipin
UT)for25cycles
<5%UT(>95%dipin
UT)for5sec
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheRadiForceseriesrequirescontinued
operationduringpowermainsinterruptions,it
isrecommendedthattheRadiForceseriesbe
poweredfromanuninterruptiblepowersupplyor
abattery.
Powerfrequency
(50/60Hz)
magneticeld
IEC/EN61000-4-8
3A/m 3A/m Powerfrequencymagneticeldsshouldbeat
levelscharacteristicofatypicallocationina
typicalcommercialorhospitalenvironment.
NOTEUTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.