44 Appendix
EMC InformationEssential performance of RadiForce series is to display images and operate functions normally.
CAUTION
The RadiForce series requires special precautions regarding EMC and need to be installed, put into
service and used according to the following information.
Do not use any cables other than the cables that provided or speci ed by us.
Using other cables may cause the increase of emission or decrease of immunity.
Do not put any portable and mobile RF communications equipment close to the RadiForce series.
Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation
in the con guration in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts,
con guring a medical system, is responsible that the system complies with the requirements of IEC/
EN60601-1-2.
Guidance and manufacturer’s declaration - electromagnetic emissions
TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.
Emission test Compliance Electromagnetic environment - guidance
RFemissions
CISPR11/EN55011
Group1 TheRadiForceseriesusesRFenergyonlyforitsinternalfunction.
Therefore,itsRFemissionareverylowandarenotlikelytocauseany
interferenceinnearbyelectronicequipment.
RFemissions
CISPR11/EN55011
ClassB TheRadiForceseriesissuitableforuseinallestablishments,includingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Harmonicemissions
IEC/EN61000-3-2
ClassD
Voltageuctuations/
ickeremissions
IEC/EN61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
TheRadiForceseriesisintendedforuseintheelectromagneticenvironmentspeciedbelow.
ThecustomerortheuseroftheRadiForceseriesshouldassurethatitisusedinsuchanenvironment.
Immunity test IEC/EN60601 test
level
Compliance level Electromagnetic environment - guidance
Electrostatic
discharge(ESD)
IEC/EN61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Electricalfast
transient/burst
IEC/EN61000-4-4
±2kV
forpowersupplylines
±1kV
forinput/outputlines
±2kV
forpowersupplylines
±1kV
forinput/outputlines
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Surge
IEC/EN61000-4-5
±1kVline(s)toline(s)
±2kVline(s)toearth
±1kVline(s)toline(s)
±2kVline(s)toearth
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.
Voltagedips,short
interruptionsand
voltagevariationson
powersupplyinput
lines
IEC/EN61000-4-11
<5%UT(>95%dipin
UT)for0.5cycle
40%UT(60%dipinUT)
for5cycles
70%UT(30%dipinUT)
for25cycles
<5%UT(>95%dipin
UT)for5sec
<5%UT(>95%dipin
UT)for0.5cycle
40%UT(60%dipin
UT)for5cycles
70%UT(30%dipin
UT)for25cycles
<5%UT(>95%dipin
UT)for5sec
Mainspowerqualityshouldbethatofatypical
commercialorhospitalenvironment.Iftheuser
oftheRadiForceseriesrequirescontinued
operationduringpowermainsinterruptions,it
isrecommendedthattheRadiForceseriesbe
poweredfromanuninterruptiblepowersupplyor
abattery.
Powerfrequency
(50/60Hz)
magneticeld
IEC/EN61000-4-8
3A/m 3A/m Powerfrequencymagneticeldsshouldbeat
levelscharacteristicofatypicallocationina
typicalcommercialorhospitalenvironment.
NOTEUTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.