37
Chapter7Reference
Environmental
conditions
Temperature Operating: 0°Cto35°C(32°Fto95°F)
Storage: -20°Cto60°C(-4°Fto140°F)
Humidity Operating: 20%to80%R.H.(nocondensation)
Storage: 10%to80%R.H.(nocondensation)
Airpressure Operating: 700to1,060hPa
Storage: 200to1,060hPa
USB Standard USBSpecicationRevision2.0
Port Upstreamport×1,Downstreamport×2
Communication
Speed
480Mbps(high),12Mbps(full),1.5Mbps(low)
Supplycurrent Downstream:Max.500mA /1port
ClassicationofEquipment Typeofprotectionagainstelectricsho ck:ClassI
EMCclass:EN60601-1-2:2007Group1ClassB
Classicationofmedicaldevice(MDD93/42/ EEC):ClassI
Main Default SettingsCAL Switch Mode : The default display mode set ting is DICOM.
Mode Brightness Temperature Gamma value
DICOM 400cd/m2(Approx.35%) 7,500K DICOM
Custom Approx.400cd/m2(35%) 7,500K 2.2
CAL1 300cd/m2(Approx.25%) 6,500K DICO M
CAL2 300cd/m2(Approx.20%) 6,500K 2.2
CAL3 400cd/m2(Approx.35%) 7,500K DICOM
sRGB Approx.250cd/m2(0%) 6,500K 2.2
Text Approx.100cd/m2(50%) 6,500K 2.2
Others
Input*DVI
PresenceSensor Off
PowerSave High
InputSelection Auto
ScreenSize Normal
Logo*Off
DDC*On
DC5VOutput*Off
SignalSelection*DualLink
MenuPosition Center
Language English
* Thesefunctionscannotbeinitializedwiththe<A llReset>function(seepage22).