Emerson 4500 manual Recall Factory Trim Recall Factory Trim- Sensor Trim

Models: 4500

1 70
Download 70 pages 45.62 Kb
Page 47
Image 47

Reference Manual

00809-0100-4027, Rev AA December 2005

Rosemount 4500

NOTE

Select pressure input values so that low and high values are equal to or outside the 4 and 20 mA points. Do not attempt to obtain reverse output by reversing the high and low points. The transmitter allows approximately five percent URL deviation from the characterized curve established at the factory.

4.Follow the commands provided by the HART Communicator to complete the adjustment of the lower value.

5.Repeat the procedure for the upper value, replacing 2: Lower sensor trim

with 3: Upper sensor trim in Step 3.

Recall Factory Trim

Recall Factory Trim— Sensor Trim

Fast Keys

1, 2, 3, 4, 1

 

 

AMS

Right click on the device and select “Calibrate,” then “Sensor trim” from the menu.

1.Select “Lower sensor trim.”

2.Click Next after setting the control loop to manual.

3.Click Next after applying appropriate pressure to sensor.

4.Select Next to acknowledge the loop can be returned to automatic control.

5.Select Finish to acknowledge the method is complete.

6.Right click on the device and select “Calibrate,” select “Sensor trim” from the menu.

7.Select “Upper sensor trim” and repeat steps 2-5.

The Recall Factory Trim commands allow the restoration of the as-shipped factory settings of the sensor trim and analog output trim.

This command resets the transmitter sensor trim to the “as shipped” factory settings. The Recall Factory Trim—Sensor Trim command can be useful for recovering from an inadvertent zero trim of an absolute pressure unit or inaccurate pressure source.

AMS

Right click on the device and select “Calibrate,” then “Recall Factory Trim” from the menu.

1.Click Next after setting the control loop to manual.

2.Select “Sensor trim” under “Trim to recall” and click Next.

3.Click Next to acknowledge restoration of trim values is complete.

4.Select Next to acknowledge the loop can be returned to automatic control.

5.Select Finish to acknowledge the method is complete.