Rosemount 644
Reference Manual
00809-0400-4728, Rev AA June 2011
Accuracy
Table A-1. Rosemount 644 Input Options and Accuracy
Sensor | Sensor | Input |
| Recommended | Digital | D/A | |||
Options | Reference | Ranges |
| Min. Span(1) | Accuracy(2) | Accuracy(3) | |||
| °C | °F |
| °C | °F | °C | °F |
| |
Pt 100 (α = 0.00385) | IEC 751 | 1562 | 10 | 18 | ± 0.15 | ± 0.27 | ±0.03% of span | ||
Pt 200 (α = 0.00385) | IEC 751 | 1562 | 10 | 18 | ± 0.15 | ± 0.27 | ±0.03% of span | ||
Pt 500 (α = 0.00385) | IEC 751 | 1562 | 10 | 18 | ± 0.19 | ± 0.34 | ±0.03% of span | ||
Pt 1000 (α = 0.00385) | IEC 751 | 10 | 18 | ± 0.19 | ± 0.34 | ±0.03% of span | |||
Pt 100 (α = 0.003916) | JIS 1604 | 10 | 18 | ± 0.15 | ± 0.27 | ±0.03% of span | |||
Pt 200 (α = 0.003916) | JIS 1604 | 10 | 18 | ± 0.27 | ± 0.49 | ±0.03% of span | |||
Ni 120 | Edison Curve No. 7 | 572 | 10 | 18 | ± 0.15 | ± 0.27 | ±0.03% of span | ||
Cu 10 | Edison Copper Winding No. 15 | 482 | 10 | 18 | ±1.40 | ± 2.52 | ±0.03% of span | ||
Pt 50 (α = 0.00391) | GOST | 1022 | 10 | 18 | ± 0.30 | ± 0.54 | ±0.03% of span | ||
Pt 100 (α = 0.00391) | GOST | 1022 | 10 | 18 | ± 0.15 | ± 0.27 | ±0.03% of span | ||
Cu 50 (α = 0.00426) | GOST | 392 | 10 | 18 | ±1.34 | ± 2.41 | ±0.03% of span | ||
Cu 50 (α = 0.00428) | GOST | 10 | 18 | ±1.34 | ± 2.41 | ±0.03% of span | |||
Cu 100 (α = 0.00426) | GOST | 392 | 10 | 18 | ±0.67 | ± 1.20 | ±0.03% of span | ||
Cu 100 (α = 0.00428) | GOST | 10 | 18 | ±0.67 | ± 1.20 | ±0.03% of span | |||
Thermocouples(4) |
|
|
|
|
|
|
|
|
|
Type B(5) | NIST Monograph 175, IEC 584 | 100 to 1820 | 212 to 3308 | 25 | 45 | ± 0.77 | ± 1.39 | ±0.03% of span | |
Type E | NIST Monograph 175, IEC 584 | 25 | 45 | ± 0.20 | ± 0.36 | ±0.03% of span | |||
Type J | NIST Monograph 175, IEC 584 | 1400 | 25 | 45 | ± 0.35 | ± 0.63 | ±0.03% of span | ||
Type K(6) | NIST Monograph 175, IEC 584 | 2501 | 25 | 45 | ± 0.50 | ± 0.90 | ±0.03% of span | ||
Type N | NIST Monograph 175, IEC 584 | 2372 | 25 | 45 | ± 0.50 | ± 0.90 | ±0.03% of span | ||
Type R | NIST Monograph 175, IEC 584 | 0 to 1768 | 32 to 3214 | 25 | 45 | ± 0.75 | ± 1.35 | ±0.03% of span | |
Type S | NIST Monograph 175, IEC 584 | 0 to 1768 | 32 to 3214 | 25 | 45 | ± 0.70 | ± 1.26 | ±0.03% of span | |
Type T | NIST Monograph 175, IEC 584 | 25 | 45 | ± 0.35 | ± 0.63 | ±0.03% of span | |||
DIN Type L | DIN 43710 | 1652 | 25 | 45 | ± 0.35 | ± 0.63 | ±0.03% of span | ||
DIN Type U | DIN 43710 | 25 | 45 | ± 0.35 | ± 0.63 | ±0.03% of span | |||
Type W5Re/W26Re | ASTM E | 0 to 2000 | 32 to 3632 | 25 | 45 | ± 0.70 | ± 1.26 | ±0.03% of span | |
GOST Type L | GOST R | 1472 | 25 | 45 | ± 1.00 | ± 1.26 | ±0.03% of span | ||
Other Input Types |
|
|
|
|
|
|
|
|
|
Millivolt Input |
| 100 mV |
|
|
| ±0.015 | mV | ±0.03% of span | |
| 0 to 2000 ohms |
|
|
| ±0.45 ohm | ±0.03% of span |
(1)No minimum or maximum span restrictions within the input ranges. Recommended minimum span will hold noise within accuracy specification with damping at zero seconds.
(2)The published digital accuracy applies over the entire sensor input range. Digital output can be accessed by HART or FOUNDATION fieldbus Communications or Rosemount control system.
(3)Total Analog accuracy is the sum of digital and D/A accuracies. This is not applicable for FOUNDATION fieldbus.
(4)Total digital accuracy for thermocouple measurement: sum of digital accuracy +0.5 °C. (cold junction accuracy).
(5)Digital accuracy for NIST Type B T/C is ±3.0 °C (±5.4 °F) from 100 to 300 °C (212 to 572 °F).
(6)Digital accuracy for NIST Type K T/C is ±0.70 °C (±1.26 °F) from
Accuracy Example
When using a Pt 100 (α = 0.00385) sensor input:
•Total accuracy = ±0.15 °C.
•No D/A accuracy effects apply