Emerson PN 51-1055pHC/rev.K instruction manual Choose Alarm Configure, Choose Measurement or Temp

Models: PN 51-1055pHC/rev.K

1 87
Download 87 pages 1.18 Kb
Page 44
Image 44

MODEL SOLU COMP II

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.4.3 Procedure: Configuring Alarms

To choose a menu item, move the cursor to the item and press ENTER.

To store a number or setting, press ENTER.

CalibrateHold

ProgramDisplay

OutputsAlarms

Measurement>>

Alarm Setpoints

Alarm Configure

1.Press MENU. The main menu screen appears. Choose Program.

2.Choose Alarms.

3.Choose Alarm Configure.

Alarm Config?

AL1 AL2 AL3

4. Choose Alarm 1 (AL1), Alarm 2 (AL2), or Alarm 3 (AL3).

AL1 is for?

Sensor1Sensor2

AL1 S1 is for?

Measurement Temp

5.For AL1 or AL2

a.Choose Sensor 1 (pH) or Sensor 2 (conductivity).

b.Choose Measurement or Temp.

c.Choose High, Low or USP 24. USP 24 appears only if Sensor 2 was selected.

d.Set the alarm Deadband.

6.The display returns to the Alarm Configure? screen. Select another alarm or press EXIT to return to the previous screen. To return to the main display, press MENU followed by EXIT.

AL3 is for?

Fault

Sensor1

Sensor2

 

 

7.For AL3

a.Choose Sensor1 (pH), Sensor2 (conductivity), or Fault.

b.For sensor 1 or 2, choose Measurement or Temp.

c.Choose High, Low or USP 24. USP 24 appears only if Sensor 2 was selected. Set the deadband.

d.Choosing Fault means AL3 will activate when a sensor or analyzer fault exists. There is no user setting to make.

8.The display returns to the Alarm Configure? screen. Select another alarm or press EXIT to return to the previous screen. To return to the main display, press MENU followed by EXIT.

35

Page 44
Image 44
Emerson PN 51-1055pHC/rev.K instruction manual Choose Alarm Configure, Choose Measurement or Temp