MODEL SOLU COMP II

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.8 NOISE REJECTION

5.8.1 Purpose.

For maximum noise rejection, the frequency of the ac power must be entered in the analyzer.

5.8.2. Procedure.

To choose a menu item, move the cursor to the item and press ENTER.

To store a number or setting, press ENTER.

CalibrateHold

ProgramDisplay

OutputsAlarms

Measurement>>

TempSecurity

#Sensors>>

Noise Rejection

ResetAnalyzer >>

1.Press MENU. The main menu screen appears. Choose Program.

2.Choose >>.

3.Choose >>.

4.Choose Noise Rejection.

5.Enter the mains frequency, 50 Hz or 60 Hz.

6.The display returns to the Noise Rejection screen. To return to the main menu, press EXIT. To return to the main display, press MENU followed by EXIT.

5.9 SINGLE SENSOR OR DUAL SENSOR INPUT

5.9.1 Purpose

The Solu Comp II accepts input from a single sensor or from two sensors. This section describes how to program the analyzer for single or dual sensors. COMPLETE THIS SECTION BEFORE DOING OTHER PROGRAMMING.

5.9.2 Procedure.

To choose a menu item, move the cursor to the item and press ENTER.

To store a number or setting, press ENTER.

CalibrateHold

ProgramDisplay

OutputsAlarms

Measurement>>

TempSecurity

#Sensors>>

#of sensors?

OneΤωο

1.Press MENU. The main menu screen appears. Choose Program.

2.Choose >>.

3.Choose #Sensors.

4.Choose One or Two. Choosing One configures the analyzer as a pH ana- lyzer. Conductivity cannot be selected for single measurement.

5.The display returns to the # Sensors screen. To return to the main menu, press MENU. To return to the main display, press MENU followed by EXIT.

42

Page 51
Image 51
Emerson PN 51-1055pHC/rev.K instruction manual Choose Noise Rejection