MODEL SOLU COMP II

SECTION 5.0

 

PROGRAMMING THE ANALYZER

5.5.4 Procedure.

To choose a menu item, move the cursor to the item and press ENTER.

To store a number or setting, press ENTER.

CalibrateHold

ProgramDisplay

OutputsAlarms

Measurement>>

Config?

Sensor1Sensor2

S1Measure?pH

RedoxORP

S1 Glass

Fault

 

Enable?

Yes

No

 

 

 

Soln Temp Corr

Sensor Isoptntl

1.Press MENU. The main menu screen appears. Choose Program.

2.Choose Measurement.

3.Choose Sensor 1 (pH) or Sensor 2 (conductivity). For a single input con- figuration, the Sensor 1 Sensor 2 screen does not appear.

If you chose Sensor 1, go to step 4. If you chose Sensor 2, go to step 11.

4.For Sensor 1, choose pH, Redox, or ORP. If you chose pH, do steps 5 through 9.

If you chose ORP or Redox, do step 10.

5.Choose NO if the pH sensing electrode is NOT a glass electrode.

6.Choose Soln Temp Corr or Sensor Isoptntl.

S1 SolnTempCorr?

Off Ultrapure >>

Sensor Isoptntl

S1: 07.00pH

S2

Measure?

Cond

TDS

 

Resistivity

 

 

 

 

 

 

S2

Temp Corr?

 

NeutSalt

Other

 

 

 

7.For Soln Temp Corr, choose Off, UltraPure, HighpH, or Custom. For Custom, enter the desired temperature coefficient.

8.For Sensor Isoptntl, enter the desired sensor isopotential pH. Do not change the sensor isopotential pH unless the sensor is known to have an isopotential pH different from 7.00.

9.The display returns to the screen shown in step 3. Choose Sensor 2 (conductivity) and follow the prompts. See step 11.

10.If Redox or ORP was selected, there are no further settings to make. The display returns to the screen shown in step 3. Choose Sensor 2 (con- ductivity) and follow the prompts. See step 11.

11.For Sensor 2, choose Cond (conductivity), Resistivity, or TDS (total dis- solved solids).

If you chose Cond or Resistivity, go to step 12. If you chose TDS, go to step 14.

12.Choose NeutSalt, Slope, Cation, or Raw. For Slope, enter the desired temperature coefficient in %/°C.

13.The display returns to the screen shown in step 3. Press EXIT to return to the previous screen. To return to the main display, press MENU followed by EXIT.

14.If TDS was selected, there are no further settings to make.The display returns to the screen shown in step 3. Press EXIT to return to the previ- ous screen. To return to the main display, press MENU followed by EXIT.