MODEL SOLU COMP II

SECTION 6.0

 

CALIBRATION

6.3 AUTO CALIBRATION — pH

6.3.1 Purpose

1.New sensors must be calibrated before use. Regular recalibration is also necessary.

2.Use auto calibration instead of manual calibration. Auto calibration avoids common pitfalls and reduces errors.

6.3.2 Definitions

1.AUTO CALIBRATION. The analyzer recognizes the buffers and uses temperature-corrected pH values in the calibration. The table lists the buffers the Solu Comp II recognizes.

 

pH at 25°C

Standard(s)

 

 

(nominal pH)

 

 

 

1.68

NIST, DIN 19266, JSI 8802, BSI (see note 1)

 

 

3.56

NIST, BSI

 

 

3.78

NIST

 

 

4.01

NIST, DIN 19266, JSI 8802, BSI

 

 

6.86

NIST, DIN 19266, JSI 8802, BSI

 

 

7.00

(see note 2)

 

 

7.41

NIST

 

 

9.18

NIST, DIN 19266, JSI 8802, BSI

 

 

10.01

NIST, JSI 8802, BSI

 

 

12.45

NIST, DIN 19266

 

Note 1: NIST is National Institute of Standards, DIN is Deutsche Institute für

Normung, JSI is Japan Standards Institute, and BSI is British Standards

Institute.

Note 2: pH 7 buffer is not a standard buffer. It is a popular commercial buffer in the United States.

The Solu Comp II also measures noise and drift and does not accept calibration data until readings are sta- ble. Calibration data will be accepted as soon as the pH reading is constant to within 0.02 units for 10 sec- onds. The stability settings cannot be changed.

2.SLOPE AND OFFSET. Once the Solu Comp II suc- cessfully completes the calibration, it calculates and displays the calibration slope and offset. The slope is reported as the slope at 25°C. Figure 6-1 defines the terms.

6.3.3 Procedure

1.Obtain two buffer solutions. Ideally the buffer pH val- ues should bracket the range of pH values to be meas- ured.

2.Remove the sensor from the process liquid. If the process and buffer temperatures are appreciably dif- ferent, place the sensor in a container of tap water at the buffer temperature. Do not start the calibration until the sensor has reached the buffer temperature. Thirty minutes is usually adequate.

FIGURE 6-1. Calibration Slope and Offset

3. Calibrate the sensor by using the procedure on the following page.

48

Page 57
Image 57
Emerson PN 51-1055pHC/rev.K instruction manual PH at 25C Standards Nominal pH, Calibration Slope and Offset

PN 51-1055pHC/rev.K specifications

The Emerson PN 51-1055pHC/rev.K is a highly advanced and reliable component designed for a wide range of applications in industrial automation and control systems. This device is known for its robust performance, exceptional durability, and integration of cutting-edge technologies, making it an ideal choice for demanding operational environments.

One of the key features of the PN 51-1055pHC/rev.K is its versatility. It is compatible with various systems and can be easily integrated into existing configurations without the need for extensive modifications. This adaptability allows users to streamline their operations and enhance overall efficiency, ensuring that the PN 51-1055pHC/rev.K is suitable for diverse industrial scenarios.

Technologically, the device incorporates advanced sensor capabilities that enable accurate monitoring and measurement of critical parameters. These sensors are designed to provide real-time data, allowing operators to make informed decisions quickly. The real-time diagnostics feature is essential for maintenance, as it helps predict potential issues before they lead to system failures, thereby reducing downtime and operational costs.

The PN 51-1055pHC/rev.K also boasts a rugged construction, built to withstand harsh industrial conditions such as extreme temperatures, vibrations, and moisture. This durability ensures a long lifespan, making it a cost-effective investment for businesses looking to optimize their operations without frequent replacements or repairs.

Another critical characteristic of the Emerson PN 51-1055pHC/rev.K is its user-friendly interface. The device is designed with ease of use in mind, featuring intuitive controls and straightforward setup processes. This accessibility enhances user experience, allowing both seasoned technicians and new operators to utilize its capabilities effectively.

Additionally, the device supports various communication protocols, enabling seamless integration with other equipment and systems. This feature is crucial for creating interconnected environments where data can be shared and analyzed across multiple platforms, fostering smarter and more efficient industrial processes.

In summary, the Emerson PN 51-1055pHC/rev.K stands out for its versatility, advanced sensor technologies, rugged design, user-friendly interface, and communication capabilities. It is a powerful tool for businesses seeking improved automation and control, helping to drive efficiency and enhance operational reliability in any industrial setting.