MODEL SOLU COMP II

CalibrateHold

ProgramDisplay

CalibrateHold

ProgramDisplay

OutputsAlarms

Measurement>>

Output Range

Output Configure

Output Range?

 

Output1

Output2

 

 

 

 

Out1 S1 Range?

 

4mA

0.00pH

 

 

Out1 S1 Range?

 

20mA

14.00pH

 

 

 

 

Output Range?

 

Output1

Output2

 

 

SECTION 4.0

DISPLAY AND OPERATION

4.3PROGRAMMING AND CALIBRATING THE SOLU COMP II - TUTORIAL

Setting up and calibrating the Solu Comp II is easy. The following tutorial describes how to move around in the programming menus. For practice, the tutorial also describes how to assign conductivity values to the 4 and 20 mA outputs for sensor 1.

1.If the MENU screen (shown at the left) is not already showing, press MENU. Calibrate is blinking, which means the cursor is on Calibrate.

2.To assign pH or conductivity values to current outputs, the Program sub-

menu must be open. Press . The cursor moves to Program (Program blinking). Press ENTER. Pressing ENTER opens the Program sub-menu.

3.The Program sub-menu permits the user to set outputs, alarms, auto- matic or manual temperature compensation, and a security code. When the sub-menu opens, Outputs is blinking, which means the cursor is on

Outputs. Press or (or any arrow key) to move the cursor around the display. Move the cursor to >> and press ENTER to cause a second screen with more program items to appear. There are three screens in the Program menu. Pressing >> and ENTER in the third screen causes the display to return to the first screen (Outputs, Alarms, Measurement).

4.For practice, assign pH values to the 4 and 20 mA outputs for sensor 1. Move the cursor to Outputs and press ENTER.

5.The screen shown at left appears. The cursor is on Output Range (blink- ing). Output range is used to assign values to the low and high current out- puts. Press ENTER.

6.The screen shown at left appears. The Solu Comp II has two outputs, out- put 1 and output 2. Move the cursor to the desired output and press ENTER. For purposes of the example, choose Output 1.

7.The screen shown at left appears. Out1 S1 in the top line means output 1 (Out1) is assigned to sensor 1 (S1). Either output can be assigned to either sensor (sensor and output assignments are made under the Output Configure menu shown in step 5). Use the Out1 S1 Range? screen to assign a pH value to the 4 mA output.

a.

Use the arrow keys to change the pH to the desired value. Press

or

 

to move the cursor from digit to digit. Press

or to increase or

 

decrease the value of the digit. Holding

or

down causes the

 

numeral to continuously scroll up or down.

 

 

 

b.

To move the decimal point, press or until the cursor is on the dec-

 

imal point. Press to move the decimal point to the right. Press

to

 

move the decimal point to the left.

 

 

 

c. Press ENTER to store the setting.

8.The screen shown at left appears. Use this screen to assign a full scale pH value to the 20 mA output. Use the arrow keys to change the pH to the desired value. Press ENTER to store the setting.

9.The screen shown at left appears. To assign values to the low and high currents for output 2, select Output 2 and follow the prompts.

10.To return to the main menu, press MENU. To return to the main display press MENU then EXIT, or press EXIT repeatedly until the main display appears. To return to the previous display press EXIT.

NOTE

To store values or settings, press ENTER before pressing EXIT.

27

Page 36
Image 36
Emerson PN 51-1055pHC/rev.K instruction manual Programming and Calibrating the Solu Comp II Tutorial

PN 51-1055pHC/rev.K specifications

The Emerson PN 51-1055pHC/rev.K is a highly advanced and reliable component designed for a wide range of applications in industrial automation and control systems. This device is known for its robust performance, exceptional durability, and integration of cutting-edge technologies, making it an ideal choice for demanding operational environments.

One of the key features of the PN 51-1055pHC/rev.K is its versatility. It is compatible with various systems and can be easily integrated into existing configurations without the need for extensive modifications. This adaptability allows users to streamline their operations and enhance overall efficiency, ensuring that the PN 51-1055pHC/rev.K is suitable for diverse industrial scenarios.

Technologically, the device incorporates advanced sensor capabilities that enable accurate monitoring and measurement of critical parameters. These sensors are designed to provide real-time data, allowing operators to make informed decisions quickly. The real-time diagnostics feature is essential for maintenance, as it helps predict potential issues before they lead to system failures, thereby reducing downtime and operational costs.

The PN 51-1055pHC/rev.K also boasts a rugged construction, built to withstand harsh industrial conditions such as extreme temperatures, vibrations, and moisture. This durability ensures a long lifespan, making it a cost-effective investment for businesses looking to optimize their operations without frequent replacements or repairs.

Another critical characteristic of the Emerson PN 51-1055pHC/rev.K is its user-friendly interface. The device is designed with ease of use in mind, featuring intuitive controls and straightforward setup processes. This accessibility enhances user experience, allowing both seasoned technicians and new operators to utilize its capabilities effectively.

Additionally, the device supports various communication protocols, enabling seamless integration with other equipment and systems. This feature is crucial for creating interconnected environments where data can be shared and analyzed across multiple platforms, fostering smarter and more efficient industrial processes.

In summary, the Emerson PN 51-1055pHC/rev.K stands out for its versatility, advanced sensor technologies, rugged design, user-friendly interface, and communication capabilities. It is a powerful tool for businesses seeking improved automation and control, helping to drive efficiency and enhance operational reliability in any industrial setting.