Emerson PN 51-1055pHC/rev.K ORP Calibration, Preparation of ORP standard solutions

Models: PN 51-1055pHC/rev.K

1 87
Download 87 pages 1.18 Kb
Page 63
Image 63

MODEL SOLU COMP II

SECTION 6.0

 

CALIBRATION

6.7 ORP CALIBRATION

6.7.1 Purpose

1.For process control, it is often important to make the measured ORP agree with the ORP of a standard solu- tion.

2.During calibration, the measured ORP is made equal to the ORP of a standard solution at a single point.

6.7.2 Preparation of ORP standard solutions

ASTM D1498-93 gives procedures for the preparation of iron (II) - iron (III) and quinhydrone ORP standards. The iron (II) - iron (III) standard is recommended. It is fairly easy to make, is not particularly hazardous, and has a shelf life of about one year. In contrast, quinhydrone standards contain toxic quinhydrone and have only an eight-hour shelf life.

Iron (II) - iron (III) standard is available from Rosemount Analytical as PN R508-16OZ. The ORP of the standard solution measured against a silver-silver chloride reference electrode is 476±20mVdc at 25°C. The redox poten- tial is -476±20mVdc at 25°C.

6.7.3 Procedure

Calibrate

Hold

Program

Display

 

 

 

 

Calibrate?

 

Sensor1

Sensor2

 

 

 

 

CalSensor1?

 

Measurement

Temp

 

 

 

 

Live

600mV

CalS1

+0000mV

 

 

1.Press MENU. The main menu screen appears. Choose Calibrate.

2.Choose Sensor1 (ORP sensor).

3.Choose Measurement.

4.The top line shows the actual ORP or redox potential (S1Live). Once the reading is stable, change the number in the second line to the desired value. Press ENTER.

5.The display returns to the screen in step 2. Choosing Sensor2 will per- mit the conductivity measurement to be calibrated.

6. To return to the main display, press MENU followed by EXIT.

54

Page 63
Image 63
Emerson PN 51-1055pHC/rev.K instruction manual ORP Calibration, Preparation of ORP standard solutions