SECTION 6.0
CALIBRATION
a.Press MENU. The main menu screen appears. Choose Program.
b.Choose Measurement.
c.Choose Sensor2 (conductivity).
d.Choose Cond.
e.Choose Other.
f.Choose Raw.
g.The display returns to the screen in step c. Return to the main display by pressing MENU followed by EXIT.
6. Make the Solu Comp II reading agree with the known conductivity.
CalibrateHold
ProgramDisplay
Calibrate?
Sensor1Sensor2
Cal Sensor2?
Measurement Temp
CalS2? InProcess
MeterCellConst
Live10.00µS/cm
CalS2 10.00µS/cm
Updated S2 Cell
Const: 1.0013/cm
Calibration
Error
a.Press MENU. The main menu screen appears. Choose Calibrate.
b.Choose Sensor2 (conductivity).
c.Choose Measurement.
d.Choose InProcess.
e.The first line is the current (or live) conductivity reading. Change the number in the second line to match the conductivity of the standard. Press ENTER. Be sure the live reading is stable before making changes.
f.If the calibration is acceptable, the updated cell constant screen appears for 5 seconds. The display returns to the screen in step b.
g.If the calibration is unacceptable (cell constant is less than 0.0001/cm or greater than 100/cm), the Calibration Error screen appears for 5 seconds. The display then returns to the screen in step e. Repeat the calibration.
58