MODEL SOLU COMP II | SECTION 4.0 |
| DISPLAY AND OPERATION |
4.4 SECURITY
4.4.1 How the Security Code Works
Use the security code to prevent accidental or unwanted changes to program settings, displays, and calibration.
Enter Security
Code000
Invalid Code
1.If a security code has been programmed, pressing MENU causes the security screen to appear.
2.Enter the
3.If the entry is correct, the main menu screen appears. If the entry is incor- rect, the Invalid Code screen appears. The Enter Security Code screen reappears after 2 seconds.
4.4.2 Bypassing the Security Code
Enter 555. The main menu will open.
4.4.3 Setting a Security Code
See Section 5.7.
4.5 USING HOLD
4.5.1 Purpose
The analyzer output is always proportional to measured pH or conductivity. To prevent unwanted alarms and improper operation of control systems or dosing pumps, place the analyzer in hold before removing the sensor for calibration and maintenance. Be sure to remove the analyzer from hold once Hold calibration is complete. During hold, both outputs remain at the last value. Once in hold, the analyzer remains there indefinitely. While in hold, the screen
shown to the left appears periodically.
4.5.2 Using the Hold Function
To choose a menu item, move the cursor to the item and press ENTER.
To store a number or setting, press ENTER.
CalibrateHold
ProgramDisplay
Hold Outputs and |
| |
Alarms? | Yes | No |
1.Press MENU. The main menu screen appears. Choose Hold.
2.The Hold Outputs and Alarms ? screen appears. Choose Yes to place the analyzer in hold. Choose No to take the analyzer out of hold.
3.The main display screen will appear.