MODEL SOLU COMP II

SECTION 6.0

 

CALIBRATION

Calibrate

Hold

Program

Display

 

 

 

 

Calibrate?

 

Sensor1

Sensor2

 

 

 

 

CalSensor1?

 

Measurement

Temp

 

 

 

 

S1

Standardize

Slope

BufferCal

 

 

 

 

S1BufferCal?

 

Auto

Manual

 

 

 

 

S1AutoCal?

 

Buffer1

Buffer2

 

 

 

 

S1Live

7.00pH

AutoBuf1

Wait

 

 

 

 

S1Live

7.00pH

AutoBuf1

7.01pH

 

 

 

 

S1AutoCal?

 

Buffer1

Buffer2

 

 

 

 

S1AutoCal?

 

Buffer1

Buffer2

 

 

 

 

S1Live

7.00pH

Buf2

Wait

 

 

 

 

S1Live

7.00pH

AutoBuf2

7.01pH

 

 

 

 

S1Offset

6mV

Slope

59.16 25°C

 

 

 

 

Calibration

 

Error

 

 

 

a.Press MENU. The main menu screen appears. Choose Calibrate.

b.Choose Sensor1 (pH sensor).

c.Choose Measurement.

d.Choose BufferCal.

e.Choose Auto.

f.Choose Buffer1.

g.Rinse the sensor with water and place it in buffer 1. Be sure the glass bulb and reference junction are completely submerged. Swirl the sensor.

h.The screen at left is displayed until the reading is stable (<0.02 pH change in 10 sec). When the reading is stable, the screen in step i appears. To bypass automatic stabilization, press ENTER at any time.

i.The top line shows the actual reading (S1Live). The analyzer also identifies the buffer and displays the nominal buffer value (buffer pH at 25°C). If the

displayed value is not correct, press or to select the correct value. The nominal value will change, for example, from 7.01 pH to 6.86 pH. Press

ENTER.

j.The screen shown at left appears.

k.Remove the sensor from buffer 1, rinse it with water, and place it in buffer 2. Swirl the sensor. Choose Buffer2.

l.The screen at left is displayed until the reading is stable (<0.02 pH change in 10 sec). When the reading is stable, the screen in step m appears. To bypass automatic stabilization, press ENTER at any time.

m.The top line shows the actual reading (S1Live). The analyzer also identifies the buffer and displays the nominal buffer value (buffer pH at 25°C). If the

displayed value is not correct, press or to select the correct value. The nominal value will change, for example, from 7.01 pH to 6.86 pH. Press ENTER to accept the nominal value.

n.If the calibration was successful, the analyzer will display the offset and slope (at 25°C). The display will return to the screen in step b. Choosing Sensor2 (conductivity sensor) will permit the conductivity measurement to be calibrated.

o.If the slope is out of range (less than 45 mV/pH or greater than 60 mV/pH), an error screen appears. The display then returns to step f. Repeat the cal- ibration.

p.To return to the main display, press MENU followed by EXIT.

49

Page 58
Image 58
Emerson PN 51-1055pHC/rev.K instruction manual Choose Measurement Choose BufferCal, Enter

PN 51-1055pHC/rev.K specifications

The Emerson PN 51-1055pHC/rev.K is a highly advanced and reliable component designed for a wide range of applications in industrial automation and control systems. This device is known for its robust performance, exceptional durability, and integration of cutting-edge technologies, making it an ideal choice for demanding operational environments.

One of the key features of the PN 51-1055pHC/rev.K is its versatility. It is compatible with various systems and can be easily integrated into existing configurations without the need for extensive modifications. This adaptability allows users to streamline their operations and enhance overall efficiency, ensuring that the PN 51-1055pHC/rev.K is suitable for diverse industrial scenarios.

Technologically, the device incorporates advanced sensor capabilities that enable accurate monitoring and measurement of critical parameters. These sensors are designed to provide real-time data, allowing operators to make informed decisions quickly. The real-time diagnostics feature is essential for maintenance, as it helps predict potential issues before they lead to system failures, thereby reducing downtime and operational costs.

The PN 51-1055pHC/rev.K also boasts a rugged construction, built to withstand harsh industrial conditions such as extreme temperatures, vibrations, and moisture. This durability ensures a long lifespan, making it a cost-effective investment for businesses looking to optimize their operations without frequent replacements or repairs.

Another critical characteristic of the Emerson PN 51-1055pHC/rev.K is its user-friendly interface. The device is designed with ease of use in mind, featuring intuitive controls and straightforward setup processes. This accessibility enhances user experience, allowing both seasoned technicians and new operators to utilize its capabilities effectively.

Additionally, the device supports various communication protocols, enabling seamless integration with other equipment and systems. This feature is crucial for creating interconnected environments where data can be shared and analyzed across multiple platforms, fostering smarter and more efficient industrial processes.

In summary, the Emerson PN 51-1055pHC/rev.K stands out for its versatility, advanced sensor technologies, rugged design, user-friendly interface, and communication capabilities. It is a powerful tool for businesses seeking improved automation and control, helping to drive efficiency and enhance operational reliability in any industrial setting.