Finisar AN-2030 specifications F i n i s a r, 2Diagnostics

Models: AN-2030

1 35
Download 35 pages 39.34 Kb
Page 17
Image 17
2Diagnostics

AN-2030: Digital Diagnostic Monitoring Interface for Optical Transceivers

1

F i n i s a r

2Diagnostics

3 2 wire serial bus address 1010001X (A2h) is used to access measurements of

4transceiver temperature, internally measured supply voltage, TX bias current, TX output

5power, received optical power, and two additional quantities to be defined in the future.

6The values are interpreted differently depending upon the option bits set at address 92.

7If bit 5 “internally calibrated” is set, the values are calibrated absolute measurements,

8which should be interpreted according to the section “Internal Calibration” below. If bit 4

9“externally calibrated” is set, the values are A/D counts, which are converted into real

10units per the subsequent section titled “External Calibration”.

11Measured parameters are reported in 16 bit data fields, i.e., two concatenated bytes.

12To guarantee coherency of the diagnostic monitoring data, the host is required to

13retrieve any multi-byte fields from the diagnostic monitoring data structure (IE: Rx Power

14MSB - byte 104 in A2h, Rx Power LSB - byte 105 in A2h) by the use of a single two-

15byte read sequence across the serial interface.

16Measurements are calibrated over specified device operating temperature and voltage

17and should be interpreted as defined below. Alarm and warning threshold values

18should be interpreted in the same manner as real time 16 bit data.

19Internal Calibration

201) Internally measured transceiver temperature. Represented as a 16 bit signed twos

21complement value in increments of 1/256 degrees Celsius, yielding a total range of –

22128°C to +128°C. Temperature measurement is valid from –40°C to +125°C with an

23accuracy of ± 3°C. The temperature sensor is located in the center of the module

24

and is typically 5 to 10 degrees ho tter than the module case. See Tables 3.12 and

253.13 below for examples of temperature format.

262) Internally measured transceiver supply voltage. Represented as a 16 bit unsigned

27integer with the voltage defined as the full 16 bit value (0 – 65535) with LSB equal to

28100 µVolt, yielding a total range of 0 to +6.55 Volts. Accuracy is ±100mV.

293) Measured TX bias current in µA. Represented as a 16 bit unsigned integer with the

30current defined as the full 16 bit value (0 – 65535) with LSB equal to 2 µA, yielding a

31total range of 0 to 131 mA. Accuracy is ± 10%. Early versions of the digital

32diagnostic standard (SFF-8472) used a scale factor of 1µA/AD Count for interpreting

33laser bias current readings. SFF-8472 later changed the scale factor to the current

34value of 2µA/AD Count. All Finisar modules using a scale factor of 2µA/AD Count

35have an ASCII “A” written in byte 56 of the ‘vendor rev’ field (see table 3.1). Legacy

36Finisar modules using a scale factor of 1µA/AD Count contain either zero or ASCII

37space (20h) or one of two place holders: “X1—“, “1A—“, in location 56.

384) Measured TX output power in mW. Represented as a 16 bit unsigned integer with

39the power defined as the full 16 bit value (0 – 65535) with LSB equal to 0.1 µW,

40yielding a total range of 0 to 6.5535 mW (~ -40 to +8.2 dBm). Data is factory

9/26/02 Revision D

Page 17

Page 17
Image 17
Finisar AN-2030 specifications F i n i s a r, 2Diagnostics