Pushing the mute button will silence the Audio alarm.

7

3.2.4CHAMBER TEMP HIGH ALARM

This alarm will activate if:

The temperature of gas leaving the humidification chamber outlet is more than 4 °C higher than the chamber set temperature for more than 20 minutes. (The chamber set temperature is the sum of the Temperature Control setting and the Chamber Control setting).

The temperature of the gas leaving the humidification chamber outlet exceeds the chamber set temperature by 10 °C or more.

This alarm causes the heater circuits to switch off and the ‘Chamber Temp High’ amber LED indicator to start flashing. The audio alarm (which can be muted) is also enabled. This alarm remains active until the high temperature condition is cleared.

3.2.5 CHAMBER TEMP LOW ALARM ACTIVATION

For Chamber Set Temperature < 35 °C

For Chamber Set Temperature 35 °C

The temperature of gas leaving the humidification

The temperature of gas leaving the humidification

chamber outlet is less than 4 °C lower than the

chamber outlet is less than 31 °C for more than 20

chamber set temperature for more than 20 minutes.

minutes.

The Chamber Temp Low alarm amber LED will begin to flash and the audio alarm is enabled. This alarm remains active until the low temperature condition is cleared.

3.2.6SET LOW WARNING

Version 1.0 and Version 2.0 Software

This alarm will activate if the chamber set temperature is below 31.0 °C. In this case the Set Low amber LED will immediately be enabled and remain on, until such time as the chamber set temperature is adjusted above 31.0 °C. There is no audible alarm.

Version 3.0 Software

This alarm will activate if either the chamber set temperature or Temperature Control setting is below 34.0 °C.

In these cases the Set Low amber LED will immediately be enabled and remain on, until such time as the set airway or chamber set temperature is adjusted above 34.0 °C. There is no audible alarm.

3.3TEMPERATURE ALARMS IN NON HEATER WIRE MODE (INCLUDES MR480)

3.3.1HEATER WIRE ALARM

This indicates that a heater wire has been connected when the non heater wire mode has been selected (Heater Wire ‘On’ LED is off). In this case the Heater Wire alarm LED will be flashing, the audio alarm begins and the heaterplate is switched off.

3.3.2AIRWAY TEMPERATURE HIGH ALARM

The red Airway Temp High LED will flash and an audible alarm will sound if the gas temperature being delivered to the patient is 41 °C or higher. Power to the heaterplate is immediately shut off.

3.3.3AIRWAY TEMPERATURE LOW ALARM

The red Airway Temp Low LED will flash and an audible alarm will sound if the gas temperature being delivered to the patient is 29.5 °C or less.

The audio alarm is not enabled for 30 seconds after the low temperature LED begins flashing.

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Page 13
Image 13
Fisher & Paykel MR730, MR720, MR480, MR700 technical manual Version 1.0 and Version 2.0 Software, Version 3.0 Software