B2 | MAINTENANCE SCHEDULE CHECKLIST |
|
|
|
|
| ||||
| (PHOTOCOPY AND FILE IN MAINTENANCE RECORDS) |
|
|
| ||||||
| Type of service check: | qmonthly | q3 monthly | q6 monthly | qyearly | |||||
| Humidifier Serial Number:___________________________ | Date: ______________ |
|
| ||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| Step |
|
|
|
| Method | Pass | Fail |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Monthly |
| Check Temperature Probe |
|
| Refer Appendix B1.1 | q | q | ||
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| Check Reusable Heater Wire |
|
| Refer Appendix B1.1 | q | q | ||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| Check Heater Wire Connections |
|
| Refer Appendix B1.1 | q | q | ||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| Check Power Cable |
|
| Refer Appendix B1.1 | q | q | ||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
| Perform |
|
| Refer Appendix B1.1 | q | q | ||
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
| ||
| Three Monthly |
| Perform Temperature Probe Test and Check |
| see § 4.3 and Appendix B | q | q | |||
|
|
| Temperature Probe |
|
|
|
|
| ||
| Annually |
| Performance Check |
|
| see § 6 | q | q | ||
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
| ||
| Electrical Safety Checks - as required |
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Method |
|
|
| Pass | Fail |
|
|
|
|
|
| |||||
|
|
|
|
|
| |||||
| Earth resistance test |
| Measure the resistance from the earth pin on the mains plug to: |
|
| |||||
|
|
|
|
| i The heaterplate.(0.2 Ω maximum) (To get a good contact on the | q | q | |||
|
|
|
|
| anodised heaterplate, clip under the front edge of it.) |
|
| |||
|
|
|
|
|
|
|
| |||
|
|
|
|
| ii The earth stud on the base of the humidifier(0.2 Ω maximum). | q | q | |||
| Insulation resistance test |
| Use a 500V DC tester to measure the resistance from the phase pin |
|
| |||||
|
|
|
|
| on the mains plug to: |
|
|
|
|
|
|
|
|
|
| i The heaterplate (10 M Ω minimum). |
| q | q | ||
|
|
|
|
| ii The earth stud on the base of the humidifier (10 M Ω minimum). | q | q | |||
| Earth Leakage test |
|
| Has to be measured using an electrical safety tester as per IEC601. | q | q | ||||
|
|
|
|
| Measure earth leakage with the earth probe of the tester on the |
|
| |||
|
|
|
|
| heaterplate. Earth leakage current is measured at normal operating |
|
| |||
|
|
|
|
| voltage and frequency of supply. |
|
|
| ||
|
|
|
|
| Should be 0.5 mA maximum in no fault condition. |
|
| |||
|
|
|
|
|
| Signature:______________________ |
|
|