16

6.3.4ALARMS CHECK - MR480

1Disconnect the 34.5°C temperature probe and check that the Temperature Probe LED flashes and gives an audio alarm.

2Insert the 41°C temperature probe into the humidifier and check that the Airway Temperature High LED flashes and the audio alarm sounds.

3Push the Mute button and check that the audio alarm is silenced

4Go to the engineering menu and check that the heaterplate duty cycle (dc) is starting to decrease, this may take a few seconds.

5Turn off the humidifier.

6Humidifier is now ready for use.

6.4DUAL PROBE TEMPERATURE TEST

This test is recommended to be performed on all temperature probes every three months or if there is any doubt as to the accuracy of the temperature probe. This test allows the operator to compare the temperature readings of both probes (chamber and airway) firstly to one another, and secondly to a glass thermometer for absolute verification of temperature if so desired. (Using a water bath and thermometer is the most reliable method to check the temperature probe accuracy.)

6.4.1EQUIPMENT REQUIRED

Dual Temperature probe to be tested.

Glass thermometer accurate to ±1°C.

Cup of warm water (Ambient to 40 °C).

Heater wire if using humidifier in heater wire mode.

6.4.2TEST PROCEDURE

1Plug the dual probe plug into the Humidifier, if in Heater Wire mode ensure a heater wire assembly is connected.

2Turn on the mains power.

3Insert both temperature probes into a container filled with water at a temperature between ambient and 40 °C. As temperature gradients can be quite large it is recommended that the two probes be held together with a rubber band during this test and the water constantly stirred.

4Place the thermometer in the water alongside the probes and allow two minutes for the temperatures to stabilise. It is recommended that the water be constantly stirred during this test.

5Enter the Engineering menu (refer Appendix A2) and select ‘CP’ (compare probes) option. The display will alternate between the airway and chamber temperatures (chamber temperature is displayed with all decimal points turned on).

6Check the temperature difference between chamber and airway probes as indicated by the display, it should not exceed 1.5 °C.

7Ensure that the temperature reading of the chamber and airway probes are both within 1.5 °C of the thermometer reading.

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Page 22
Image 22
Fisher & Paykel MR480, MR720, MR730, MR700 technical manual Dual Probe Temperature Test