| Excessive water in expiratory | Condensation due to cooling in | Use dual heated circuit. |
| circuit due to rainout. | Expiratory line. |
|
|
|
|
|
| Water ejected from | Humidification chamber filled beyond | Rduce water level |
| humidification chamber into | maximum water level line. |
|
| circuit. |
|
|
|
|
|
|
|
| Maximum permissible peak flow rate of | Ue correct chamber. Refer to the |
|
| humidification chamber has been | operating instructions for the type of |
|
| exceeded. | chamber used. |
|
|
|
|
|
| Humidifier chamber tilted. | Chamber should be level. |
|
|
|
|
C1.5 | LOW HUMIDITY |
|
|
|
|
|
|
| Condition detected | Possible cause | FAction to take |
|
|
|
|
|
|
|
|
| Low humidity. | Chamber control set to low. | Increase setting. |
|
|
|
|
|
| No water in humidification chamber. | Fill chamber. |
|
|
|
|
|
| Chamber temperature low. | Check heaterplate temperature setting on |
|
|
| diaplay when humidifier is switched on, |
|
|
| Chamber base and heaterplate are smooth |
|
|
| and flat and in good contact. (Chambers |
|
|
| not made by Fisher & Paykel may not be |
|
|
| suitable). |
|
|
|
|
|
| Gas flow rate too high for chamber. | Check maximum flow rate in to chamber |
|
|
| operating instructions. |
|
|
|
|
C1.6 | ALARMS |
|
|
|
|
|
|
| Condition detected | Possible cause | FAction to take |
|
|
|
|
|
|
|
|
| Temperature probe alarm | Temperature probe unplugged. | Check connetion. |
| appears on display). |
|
|
|
|
|
|
|
| Temperature probe plug dirty. | Clean plug. |
|
|
|
|
|
| Temperature probe faulty | Replace probe. |
|
|
|
|
| Heater wire alarm while | Heater wire unplugged. | Check connection. |
| temperature still displayed. |
|
|
|
|
|
|
|
| Heater wire broken. | Replace heater wire. |
|
|
|
|
|
| Heater wire switch turned off but | Turn heater wire on. |
|
| heater wire connected. |
|
|
|
|
|
|
| Heater wire switch turned on but heater | Turn heater wire off. |
|
| wire not connected. |
|
|
|
|
|
| Heater wire alarm while ‘F2’ | Internal fuse blown. | Send for repair. |
| displayed. |
|
|
|
|
|
|
|
| 41 °C Overheat calibration outside | Check calibration of the 41 °C |
|
| limits. | comparator. |
|
|
|
|