Fisher & Paykel MR700, MR720, MR730, MR480 technical manual LED on

Models: MR700 MR480 MR720 MR730

1 64
Download 64 pages 47.74 Kb
Page 59
Image 59

 

 

53

 

C1.7

TEMPERATURE DISPLAY

 

 

 

 

 

 

 

Condition detected

Possible cause

FAction to take

 

 

 

 

 

 

 

 

 

No temperature display, no alarm

Humidifier voltage rating incorrect.

Select correct voltage rated unit.

 

lights.

 

 

 

 

 

 

 

 

Heaterplate overheat protector

Reset overheat protector.

 

 

operated.

 

 

 

 

 

 

 

Transformer over heat protector

Replace transformer.

 

 

operated.

 

 

 

 

 

 

No temperature display, ‘see

Control PCB fault.

Check control PCB for water damage.or

 

manual’ indicator light on.

 

contamination.

 

 

 

 

 

Error code.

Humdifier failed self test.

Check control PCB for water damage or

 

 

 

contamination.

 

 

 

Check thermistor conection to control

 

 

 

PCB.

 

 

 

If new PCB fitted, check correct polarity

 

 

 

of transformer secondary harness.

 

 

 

 

C1.8

BREATHING CIRCUIT LEAK

 

 

 

 

 

 

 

Condition detected

Possible cause

FAction to take

 

 

 

 

 

 

 

 

 

Breathing circuit leak.

Faulty circuit connections.

Check all circuit connections.

 

 

 

 

 

 

No water bag.

Connect water bag

 

 

 

 

 

 

Temperature probes incorrectly fitted.

Check connections.

 

 

 

 

 

 

Chamber filler stopper missing.

Put stopper in place.

 

 

 

 

C1.9

EXTERNAL WATER LEAK

 

 

 

 

 

 

 

Condition detected

Possible cause

FAction to take

 

 

 

 

 

 

 

 

 

Water leak with reusable

Chamber O-ring faulty.

O-ring should be clean and properly

 

chamber.

 

fitted. Replace if damaged.

 

 

 

 

 

Water leak with disposable

Chamber used beyond lifespan.

Replace with new chamber.

 

chamber.

 

 

 

 

 

 

C1.10

LED ON

 

 

 

 

 

 

 

Condition detected

Possible cause

FAction to take

 

 

 

 

 

 

 

 

 

An alarm LED is flashing.

No audible sound - mute switch has

Wait three minutes.

 

 

been pressed.

 

 

 

 

 

 

Standby LED is flashing.

Humidifier is in standby mode.

Puch button for 1 second to disable.

 

 

 

Wait 20 minutes (30 minutes for v3.0

 

 

 

software).

 

 

 

 

 

‘Set Low’ LED is on.

Chamber control or airway temperature

Re-adjust. Refer to humidifier operating

 

 

control setting too low.

manual.

 

 

 

 