51

 

C.1.2

HIGH TEMPERATURE

 

 

 

 

 

 

 

Condition detected

Possible cause

F Action to take

 

 

 

 

 

 

 

 

 

Airway temperature high alarm

Sudden spurious increase in gas flow.

Wait for temperature to fall below

 

activated.

 

threshold.

 

 

 

 

 

 

Heater wire too close to temperature

Use shorter heater wire or lengthen

 

 

probe.

inspiratory hose so that distance is

 

 

 

between 25 (1”) and 100 mm (4”), ideally

 

 

 

30 mm.

 

 

 

 

 

 

Possibly faulty EOH sensor on

Check probe, see § 6.4.

 

 

temperature probe (low resistance).

 

 

 

 

 

 

Chamber temperature high alarm

No water in chamber.

Refill humidification chamber.

 

activated.

 

 

 

 

 

 

 

 

Possibly faulty TOP sensor on

Check probe, see § 6.4.

 

 

temperature probe (low resistance).

 

 

 

 

 

C.1.3

TEMPERATURE FLUCTUATING

 

 

 

 

 

 

 

Condition detected

Possible cause

FAction to take

 

 

 

 

 

 

 

 

 

High and low temperature

Ventilation rate too low or too erratic

Ventilation rates below about 6 bpm or of

 

alarms.

for proper humidifier temperature

a highly erratic nature may affect the

 

 

control.

stability of the temperature control.

 

 

 

 

 

 

Humidifier temperature control

Shield breathing circuit from external

 

 

influenced by external heating or

influence.

 

 

cooling such as air conditioner.

 

 

 

 

 

 

 

Temperature probes incorrectly

Check temperature probe position in the

 

 

positioned.

circuit.

 

 

 

 

 

 

Possibly faulty temperature probe

Check probe, see Section 6.4.

 

 

(Could be intermittent short or open

(Wriggle probe to pick up intermittent

 

 

circuit).

 

 

faults).

 

 

 

 

 

 

 

C.1.4

EXCESSIVE WATER IN CIRCUIT

 

 

 

 

 

 

 

Condition detected

Possible cause

FAction to take

 

 

 

 

 

 

 

 

 

Excessive water in inspiratory

If in non-heated wire mode, water trap

Reposition water trap.

 

circuit due to rainout.

not at the lowest point of the circuit.

 

 

 

 

 

 

 

If in heated wire mode, heater wire too

Increase distance from heater wire to

 

 

close to temperature probe.

temperature probe so that the distance is

 

 

 

between 25 (1”) and 100mm (4”), ideally

 

 

 

30mm.

 

 

 

 

 

 

Inspiratory circuit being cooled by air

Shield circuit from air conditioner or use

 

 

conditioner.

heater wire circuits.

 

 

 

 

 

 

Chamber control set too high

Reduce setting.

 

 

 

 

 

 

Maximum permissible peak flow rate of

Use correct chamber. Refer to the

 

 

humidification chamber has been

operating instructions for the type of

 

 

exceeded.

chamber used.

 

 

 

 

 

 

Humidifier chamber tilted.

Ensure chamber is level.

 

 

 

 

 

 

Possibly faulty temperature probe (low

Check probe, see § 6.4

 

 

resistance).

 

 

 

 

 

 

 

Inspiratory heater wire not connected

Connect inspiratory heater wire.

 

 

in a parallel wired dual circuit.

 

 

 

 

 

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001

Page 57
Image 57
Fisher & Paykel MR730, MR720, MR480, MR700 technical manual High Temperature