Fisher & Paykel MR730, MR720, MR480, MR700 technical manual Appendix a, Off

Models: MR700 MR480 MR720 MR730

1 64
Download 64 pages 47.74 Kb
Page 49
Image 49

43

APPENDIX A ___________________________________________________

A1 CALIBRATION MODE

Entry to the calibration mode is made by holding down the ‘Chamber Temp’ button and turning on the Mains Power switch.

After a two second delay the display should show ‘-1-’.

On releasing the ‘Chamber Temp’ button the display will show the humidifier model number (730/720/700 or 480).

The ‘Chamber Temp’ button is now used to step between functions as shown below.

1Model number display.

2Software version display.

3Traceability number.

441.0 °C comparator calibrate. VR4.

5ADC zero calibrate. VR5.

On pushing the ‘Chamber Temp’ button the display will show the present function number then after a one second delay step to the next higher function number, where if the ‘Chamber Temp’ button is then released then that function is entered

The calibration mode will time out (to normal operating mode) 5 minutes after the last ‘Chamber Temp’ button push. The normal exit mode is to turn the humidifier mains power switch off then on again.

For details on 41.0 °C and ADC zero calibration procedures refer § 5.3.

A2 ENGINEERING MENU

To enter the engineering menu push the Mute and ‘Chamber Temp’ buttons simultaneously and hold for at least 1.5 seconds. At this time the display will show SET (for set temperature display) as a pointer to that function and on release of the ‘Chamber Temp’ button the display will show the set Temperature Control temperature setting to one decimal place. To maintain this display, the Mute button should be kept depressed. If the Mute is released, the display will revert to normal operation in about 6 seconds.

The display can be stepped through the engineering menu in two ways.

If the Mute is held down constantly the functions can be selected sequentially by pressing the ‘Chamber Temp’ until the display changes to the next function pointer, where if the ‘Chamber Temp’ button is then released the display will show that function.

Releasing the mute button the humidifier will automatically step through the menu until either the mute button is pressed whereupon the display will show that function or if 6 seconds has elapsed and the humidifier will display ‘End’.

If ‘End’ appears on the display and the Mute button is momentarily pushed the time out will be inhibited and the stepping through the function pointers continued.

A2.1 ENGINEERING DISPLAY FUNCTIONS: HEATER WIRE MODE

SET

Set temperature display (°C).

OFF

Offset temperature display (Chamber Control) (°C).

Edc

Heater wire duty cycle (%).

Cdc

Heaterplate duty cycle (%).

hP

Heaterplate temperature (°C).

L

Programming link status.(L1, L2, ... L6).

CP

Compare probes (Probe test).

Ei

Heater wire integral.

Ti

Heaterplate integral.

Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001