C1 TROUBLESHOOTING GUIDE FOR THE MR700 SERIES RESPIRATORY HUMIDIFIERS

This is a guide to possible corrective action in the case of an humidifier alarm condition or malfunction. It is primarily written for humidifiers using heated wire mode. So parts of this guide would not be applicable for the MR480 and MR730 and MR720 in the heater wire off mode.

Unless the cause and immediate effects of an alarm condition are understood and assessed to be of no hazard to the patient, the humidifier should be immediately switched off and removed from the patient circuit.

C1.1 LOW TEMPERATURE

Condition detected

Possible cause

F Action to take

 

 

 

 

 

 

Temperature less than desired,

Humidifier warming up.

Wait for a few minutes.

no alarm.

 

 

 

 

 

 

Standby mode switched on.

Push mode button for 1 second to

 

 

disable standby. Wait 20 minutes (30

 

 

minutes for v3.0 software).

 

 

 

 

Temperature probe not in position.

Check position of temperature probe in

 

 

the circuit.

 

 

 

Airway temperature low alarm

Gas flow stopped or interrupted.

Enter standby mode or turn humidifier

activated.

 

off until gas flow resumed. (Gas flow

 

 

must be resumed before coming out of

 

 

standby mode).

 

 

 

 

Set temperature increased by more than

Wait for temperature to rise.

 

2 °C.

 

 

 

 

 

Temperature probe not in probe

Fit temperature probe.

 

housing.

 

 

 

 

 

Gas flow rate too low.

Increase gas flow to more than 2 litres

 

 

per minute.

 

 

 

 

Too much distance between end of

Use longer heater wire or shorten

 

heater wire and temperature probe.

inspiratory hose so that distance is

 

 

between 25 (1") and 100 mm (4"), ideally

 

 

30 mm.

 

 

 

Chamber temperature low alarm

Poor thermal contact between base of

Check that chamber base and heaterplate

activated.

chamber and heaterplate.

are clean, smooth and flat. If necessary

 

 

clean or replace. Humidification

 

 

chambers which are not manufactured by

 

 

Fisher & Paykel may not be suitable.

 

 

 

 

Temperature probe not inserted at

Fit temperature probe correctly.

 

chamber outlet.

 

 

 

 

 

Heater element needs replacement.

Send for repair.

 

 

 

 

Gas flow rate too high for chamber

Use correct chamber. Refer to the

 

 

operating instructions for the type of

 

 

chamber used.

 

 

 