Digital Multimeter

Specifications

664 to 600 V Overvoltage Category IV. UL listed to UL61010-1. Licensed by TÜV to EN61010-1.

Detailed Specifications

For all detailed specifications:

Accuracy is given as ±([% of reading] + [number of least significant digits]) at 18° C to 28° C, with relative humidity up to 90 %, for a period of one year after calibration.

For Model 87 in the 4 ½-digit mode, multiply the number of least significant digits (counts) by 10. AC conversions are ac-coupled and valid from 3 % to 100 % of range. Model 87 is true rms responding. AC crest factor can be up to 3 at full scale, 6 at half scale. For non-sinusoidal wave forms add -(2 % Rdg + 2 % full scale) typical, for a crest factor up to 3.

Table 2. Model 87 AC Voltage Function Specifications

Function

Range

Resolution

Accuracy

 

 

 

 

 

 

 

 

 

 

45 - 65 Hz

30 - 200 Hz

200 - 440 Hz

440 Hz -

1 - 5 kHz

 

5 - 20 kHz1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1 kHz

 

 

 

 

 

K2,4

 

600.0 mV

0.1 mV

± (0.7 % + 4)

 

 

 

 

 

± (2.0 % + 4)

 

± (2.0 % + 20)

 

 

 

6.000 V

0.001 V

 

 

 

 

 

 

 

 

 

 

 

 

 

 

60.00 V

0.01 V

± (0.7 % + 2)

± (1.0 % + 4)

 

 

 

 

 

 

 

 

600.0 V

0.1 V

 

 

 

 

 

 

 

 

 

 

± (2.0 % + 4)3

 

unspecified

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1000 V

1 V

 

 

 

 

 

 

 

 

 

 

unspecified

 

unspecified

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Low pass filter

 

 

 

 

 

± (0.7 % + 2)

± (1.0 % + 4)

 

+1 % + 4

unspecified

unspecified

 

unspecified

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-6 % - 45

 

 

 

 

 

1.

Below 10 % of range, add 12 counts.

 

 

 

 

 

 

 

 

 

2.

The Meter is a true rms responding meter. When the input leads are shorted together in the ac functions, the Meter

 

 

may display a residual reading between 1 and 30 counts. A 30 count residual reading will cause only a 2-digit

 

 

change for readings over 3 % of range. Using REL to offset this reading may produce a much larger constant error

 

 

in later measurements.

 

 

 

 

 

 

 

 

 

 

 

3.

Frequency range: 1 kHz to 2.5 kHz.

 

 

 

 

 

 

 

 

 

4.

A residual reading of up to 13 digits with leads shorted, will not affect stated accuracy above 3 % of range.

5.

Specification increases from -1% at 200 Hz to -6% at 440 Hz when filter is in use.

 

 

 

 

 

 

 

 

Table 3. Models 83 and 88 AC Voltage Function Specifications

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Function

Range

 

 

Resolution

 

 

 

 

 

Accuracy

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

50 Hz - 60 Hz

 

30 Hz - 1 kHz

 

1 kHz - 5 kHz

 

K1

 

 

600.0 mV

 

0.1 mV

 

 

± (0.5 % + 4)

 

± (1.0 % + 4)

 

± (2.0 % + 4)

 

 

 

 

6.000 V

 

 

0.001 V

 

± (0.5 % + 2)

 

± (1.0 % + 4)

 

± (2.0 % + 4)

 

 

 

 

60.00 V

 

 

0.01 V

 

 

± (0.5 % + 2)

 

± (1.0 % + 4)

 

± (2.0 % + 4)

 

 

 

 

600.0 V

 

 

0.1 V

 

 

± (0.5 % + 2)

 

± (1.0 % + 4)

 

± (2.0 % + 4)2

 

 

 

 

1000 V

 

 

1 V

 

 

± (0.5 % + 2)

 

± (1.0 % + 4)

 

unspecified

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.

Below a reading of 200 counts, add 10 counts.

 

 

 

 

 

 

 

 

2.

Frequency range:

1 kHz to 2.5 kHz.

 

 

 

 

 

 

 

 

 

 

 

For models 83 and 88, ac conversions are ac-coupled and are average- responding, rms-indicating.

 

5

Page 13
Image 13
Fluke 80 Series V manual Detailed Specifications, Model 87 AC Voltage Function Specifications