80 Series V
Calibration Manual
Notes
Set the calibrator to Standby prior to changing the function switch position and or after completing adjustment of each function.
If the calibration adjustment procedure is not completed correctly, the Meter will not operate correctly.
Table 16. Calibration Adjustment Steps
Function |
| Adjustment |
| Input Value |
| ||
Input Terminal | Fluke |
|
| ||||
(Switch Position) | Step | Fluke | |||||
| |||||||
|
|
|
|
| |||
K | I | 600.0 mV, 60 Hz | 600.0 mV, 60 Hz | ||||
|
|
|
|
| |||
(AC Volts) |
| 600.0 mV, 5 kHz | 600.0 mV, 20 kHz | ||||
|
| 6.000 | V, 60 Hz | 6.000 | V, 60 Hz | ||
|
| 6.000 | V, 5 kHz | 6.000 | V, 20 kHz | ||
|
| 60.00 | V, 60 Hz | 60.00 | V, 60 Hz | ||
|
| 60.00 | V, 5 kHz | 60.00 | V, 20 kHz | ||
|
| 600.0 | V, 60 Hz | 600.0 | V, 60 Hz | ||
|
| 600.0 | V, 5 kHz | 600.0 | V, 10 kHz | ||
L |
| 6.000 | V | 6.000 | V | ||
|
|
|
|
|
| ||
(DC Volts) |
| 60.00 | V | 60.00 | V | ||
|
| 600.0 | V | 600.0 | V | ||
mL |
| 600.0 mV | 600.0 mV | ||||
|
|
|
|
|
| ||
(DC Millivolts) |
| 60.00 mV | 60.00 mV | ||||
e |
| 600.0 |
| 600.0 |
| ||
|
|
|
|
|
| ||
(Ohms) |
| 6.000 k | 6.000 k | ||||
|
| 60.00 k | 60.00 k | ||||
|
| 600.0 k | 600.0 k | ||||
|
| 6.000 | M | 6.000 | M | ||
|
| 0.000 |
| 0.000 |
| ||
|
| 50.0 M | 50.0 M | ||||
O |
| 3.000 | V | 3.000 | V | ||
(Diode Test) |
| ||||||
|
|
|
|
|
| ||
A | A | 6.000 | A, 60 Hz | 6.000 | A, 60 Hz | ||
6.000 | A dc | 6.000 | A dc | ||||
(Amps) |
| ||||||
mA | mA /UA | 60.00 mA, 60 Hz | 60.00 mA, 60 Hz | ||||
|
|
|
|
| |||
(Milliamps) |
| 400.0 mA, 60 Hz | 400.0 mA, 60 Hz | ||||
|
| 60.00 mA dc | 60.00 mA dc | ||||
|
| 400.0 mA dc | 400.0 mA dc | ||||
µA |
| 600.0 | ∝A ac, | 600.0 | ∝A ac, 60 Hz | ||
| 60 Hz |
| |||||
(Microamps) |
|
|
|
|
| ||
| 6000 ∝A, 60 Hz | 6000 ∝A, 60 Hz | |||||
|
| ||||||
|
| 600.0 | ∝A dc | 600.0 | ∝A dc | ||
|
| 6000 ∝A dc | 6000 ∝A dc |
26