80 Series V

Calibration Manual

Notes

Set the calibrator to Standby prior to changing the function switch position and or after completing adjustment of each function.

If the calibration adjustment procedure is not completed correctly, the Meter will not operate correctly.

Table 16. Calibration Adjustment Steps

Function

 

Adjustment

 

Input Value

 

Input Terminal

Fluke 83-V and

 

 

(Switch Position)

Step

Fluke 87-V

 

88-V

 

 

 

 

 

K

I

C-01

600.0 mV, 60 Hz

600.0 mV, 60 Hz

 

 

 

 

 

(AC Volts)

 

C-02

600.0 mV, 5 kHz

600.0 mV, 20 kHz

 

 

C-03

6.000

V, 60 Hz

6.000

V, 60 Hz

 

 

C-04

6.000

V, 5 kHz

6.000

V, 20 kHz

 

 

C-05

60.00

V, 60 Hz

60.00

V, 60 Hz

 

 

C-06

60.00

V, 5 kHz

60.00

V, 20 kHz

 

 

C-07

600.0

V, 60 Hz

600.0

V, 60 Hz

 

 

C-08

600.0

V, 5 kHz

600.0

V, 10 kHz

L

 

C-09

6.000

V

6.000

V

 

 

 

 

 

 

(DC Volts)

 

C-10

60.00

V

60.00

V

 

 

C-11

600.0

V

600.0

V

mL

 

C-12

600.0 mV

600.0 mV

 

 

 

 

 

 

(DC Millivolts)

 

C-13

60.00 mV

60.00 mV

e

 

C-14

600.0

 

600.0

 

 

 

 

 

 

 

(Ohms)

 

C-15

6.000 k

6.000 k

 

 

C-16

60.00 k

60.00 k

 

 

C-17

600.0 k

600.0 k

 

 

C-18

6.000

M

6.000

M

 

 

C-19

0.000

 

0.000

 

 

 

C-20

50.0 M

50.0 M

O

 

C-21

3.000

V

3.000

V

(Diode Test)

 

 

 

 

 

 

 

A

A

C-22

6.000

A, 60 Hz

6.000

A, 60 Hz

C-23

6.000

A dc

6.000

A dc

(Amps)

 

mA

mA /UA

C-24

60.00 mA, 60 Hz

60.00 mA, 60 Hz

 

 

 

 

 

(Milliamps)

 

C-25

400.0 mA, 60 Hz

400.0 mA, 60 Hz

 

 

C-26

60.00 mA dc

60.00 mA dc

 

 

C-27

400.0 mA dc

400.0 mA dc

µA

 

C-28

600.0

A ac,

600.0

A ac, 60 Hz

 

60 Hz

 

(Microamps)

 

 

 

 

 

 

C-29

6000 A, 60 Hz

6000 A, 60 Hz

 

 

 

 

C-30

600.0

A dc

600.0

A dc

 

 

C-31

6000 A dc

6000 A dc

26

Page 34
Image 34