Table 14. Accuracy Tests (cont.)

Step

Test

Range

5500A Output

 

Display Reading

 

 

 

Function

 

83 and 88

87

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

23

e

600

330 ( Use 2 wire Comp)1

328.5 to 331.5

329.1 to 330.9

 

Ohms

 

 

 

 

 

24

6 k

3.3 k(Use 2 wire Comp)

1

3.286 to 3.314

3.292 to 3.308

 

 

 

 

 

 

 

 

 

25

 

60 k

33 k

 

32.86 to 33.14

32.92 to 33.08

 

 

 

 

 

 

 

26

 

600 k

330 k

 

327.6 to 332.4

327.9 to 332.1

 

 

 

 

 

 

 

27

 

6 M

3.3 M

 

3.276 to 3.324

3.279 to 3.321

 

 

 

 

 

 

 

28

 

50 M

30 M

 

29.67 to 30.33

29.67 to 30.33

 

 

 

 

 

 

 

29

nS

60 nS

Open input

 

- 0.30 to 0.30

- 0.30 to 0.30

 

Conductance

 

 

 

 

 

30

60 nS

100 M

 

9.60 to 10.40

9.60 to 10.40

 

 

 

 

 

 

 

31

G

6 V

3.0 V dc

 

2.939 to 3.061

2.939 to 3.061

 

Diode

 

 

 

 

 

 

 

 

 

 

 

 

32

?

6 A

3.0 A, 60 Hz

 

2.962 to 3.038

2.968 to 3.032

 

AC Amps

 

 

 

 

 

 

 

 

 

 

 

 

33

A

6 A

3.0 A

 

2.984 to 3.016

2.990 to 3.010

 

DC Amps

 

 

 

 

 

 

 

 

 

 

 

 

33B

A

10A6

10A

 

9.94 to 10.06

9.96 to 10.04

 

DC Amps

 

 

 

 

 

 

 

 

 

 

 

 

34

m?

60 mA

33 mA, 60 Hz

 

32.58 to 33.42

32.65 to 33.35

 

 

 

 

 

 

 

 

 

35

AC Milliamps

400 mA

330 mA, 60 Hz

 

325.8 to 334.2

326.5 to 333.5

 

 

 

 

 

 

 

36

m[

60 mA

33 mA

 

32.83 to 33.17

32.89 to 33.11

 

DC Milliamp

 

 

 

 

 

37

400 mA

330 mA

 

328.5 to 331.5

329.1 to 330.9

 

 

 

 

 

 

 

38

U?

600 A

330 A, 60 Hz

 

325.8 to 334.2

326.5 to 333.5

 

 

 

 

 

 

39

6000 A

3300 A, 60 Hz

 

3258 to 3342

3265 to 3335

AC

 

 

 

 

 

 

 

 

Microamps

 

 

 

 

 

 

 

 

 

 

 

 

40

U[

600 A

330 A

 

328.3 to 331.7

328.9 to 331.1

 

DC

 

 

 

 

 

41

6000 A

3300 A

 

3285 to 3315

3291 to 3309

 

Microamps

 

 

 

 

 

 

 

 

 

 

 

 

42

E

10 nf

Open input2

 

0.21 to 0.31

0.21 to 0.31

 

Capacitance

 

 

 

 

 

43

100 nf

5 nf5

 

04.7 to 05.3

04.7 to 05.3

 

 

 

 

 

 

 

44

 

100 f

9.5 f

 

09.2 to 09.8

09.2 to 09.8

 

 

 

 

 

 

 

45

K

1000 V

400 V, 400 Hz

 

N/A

376 to 408

46

Low Pass

1000 V

400 V, 800 Hz4

 

N/A

226 to 3404

 

Filter

 

 

 

 

 

 

 

 

 

 

 

 

18

Page 26
Image 26
Fluke 80 Series V manual 600 Ω 330 Ω Use 2 wire Comp 328.5 to 329.1 to Ohms