Digital Multimeter

Required Equipment

Good F2 fuse: 00.0 to 00.5

Replace fuse: OL

Touch top half

of input contacts

Good F1 fuse: 0.995 kto 1.005 k

Replace fuse: OL

87 V TRUE RMS MULTIMETER

 

˚C/˚F

 

MIN MAX

RANGE

AutoHOLD

Peak MIN MAX

 

 

 

REL

Hz %

4½ DIGITS

 

 

1 Second

 

 

mV

 

 

 

 

mA

V

 

A

V

 

A

OFF

 

 

AmA A COM V

400mA

10A MAX FUSED FUSED

87 V TRUE RMS MULTIMETER

 

˚C/˚F

 

MIN MAX

RANGE

AutoHOLD

Peak MIN MAX

 

 

 

REL

Hz %

4½ DIGITS

 

 

1 Second

 

 

mV

 

 

 

 

mA

V

 

A

V

 

A

OFF

 

 

AmA A COM V

400mA

10A MAX FUSED FUSED

aom5f.eps

Figure 3. Testing the Current Input Fuses

Replacing the Fuses

To replace the fuse(s), perform the following procedure.

1.To open the Meter, refer to “Opening the Meter Case”. See Figure 1.

2.Grasp the fuse in the center with needle nose pliers. Pull straight up on the fuse to remove it from the fuse clips.

3.Install ONLY specified replacement fuses with the amperage, voltage, and speed ratings shown in Table 17.

4.To close the Meter, refer to “Reassembling the Meter Case”.

Required Equipment

Required equipment for the performance tests is listed in Table 13. If the recommended models are not available, equipment with equivalent specifications may be used.

XW Warning

To avoid shock or injury, do not perform the verification tests or calibration adjustment procedures described in this manual unless you are qualified to do so.

Repairs or servicing should be performed only by qualified personnel.

13

Page 21
Image 21
Fluke 80 Series V manual Required Equipment, Replacing the Fuses