Digital Multimeter

Performance Tests

Testing the Pushbuttons

To test the pushbuttons

1.Turn the Meter rotary knob to J.

2.Press each button and note that the meter responds with a beep for each button press.

3.Press and hold Ba second time to exit MIN MAX mode.

Testing Meter Accuracy

Perform the accuracy test steps in Table 15.

Table 15. Accuracy Tests

Step

Test

Range

5500A Output

Display Reading

 

Function

83 and 88

 

87

 

 

 

 

1

 

K

 

600 mV

330 mV, 60 Hz

327.9 to 332.1

327.3 to 332.7

2

 

 

600 mV

600 mV, 13 kHz

N/A

586.0 to 614.0

 

AC Volts

 

3

 

 

6 V

3.3 V, 60 Hz

3.281 to 3.319

3.275 to 3.325

 

 

 

4

 

 

 

6 V

3.3 V, 20 kHz

N/A

3.214 to 3.386

5

 

 

 

60 V

33 V, 60 Hz

32.81 to 33.19

32.75 to 33.25

6

 

 

 

60 V

33 V, 20 kHz

N/A

32.14 to 33.86

7

 

 

 

600 V

330 V, 60 Hz

328.1 to 331.9

327.5 to 332.5

8

 

 

 

600 V

330 V, 2.5 kHz

N/A

323.0 to 337.0

9

 

 

 

1000 V

500 V, 60 Hz

495 to 505

494 to 506

10

 

 

 

1000 V

1000 V, 1 kHz

986 to 1014

986 to 1014

11

 

K Hz

 

600 mV

150 mV, 99.95 kHz

99.93 to 99.97

99.93 to 99.97

12

 

 

600 mV

150 mV, 199.50 kHz

199.48 to 199.52

199.48 to 199.52

 

AC Volts

 

 

 

 

 

 

 

 

 

 

Frequency

 

 

 

 

 

13

 

Sensitivity

 

6 V

0.7 V, 99.95 kHz

99.93 to 99.97

99.93 to 99.97

14

 

 

 

60 V

7 V, 99.95 kHz

99.93 to 99.97

99.93 to 99.97

15

 

VHz

 

6 V

3.4 V, 1 kHz Sq. Wave

999.8 to 1000.2

999.8 to 1000.2

 

 

Trigger level

 

 

 

 

 

16

 

VHz

 

6 V

5 V, 1 kHz, DC offset 2.5 V

49.7% to 50.3 %

49.7 % to 50.3 %

 

 

Duty Cycle

 

 

Sq. Wave

 

 

 

 

 

 

 

 

 

17

 

V

 

6V

3.3 V dc

3.296 to 3.304

3.297 to 3.303

18

 

 

60 V

33 V dc

32.96 to 33.04

32.97 to 33.03

 

DC Volts

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

19

 

 

 

600 V

330 V dc

329.6 to 330.4

329.7 to 330.3

 

 

 

 

 

 

 

 

20

 

 

 

1000 V

1000 V dc

998 to 1002

998 to 1002

 

 

 

 

 

 

 

 

21

 

mV

 

600 mV

33 mV dc

32.8 to 33.2

32.9 to 33.1

22

 

DC Volts

 

600 mV

330 mV dc

328.9 to 331.1

329.6 to 330.4

17

Page 25
Image 25
Fluke 80 Series V manual Testing the Pushbuttons, Testing Meter Accuracy, Accuracy Tests