80 Series V
Calibration Manual
Restoring the Default Password
If the calibration password is forgotten, the default password (1234) can be restored using the following steps.
1.While holding down B, turn the rotary knob from OFF to VAC. The Meter displays “ZCAL”.
2.Remove the Meter's top case. Leave the PCA in the bottom case. (See “Opening the Meter Case”.)
XWWarning
To avoid electrical shock or personal injury, remove the test leads and any input signal before removing the Meter's top case.
3.Through an access hole provided in the top shield, short across the keypads on the PCA. See Figure 7. The Meter should beep. The default password is now restored.
4.Replace the Meter's top case and turn the rotary knob to OFF. (See “Reassembling the Meter Case).
Access Hole For
PCA Keypad
ama01f.eps
Figure 7. Restoring the Default Password
24