Meter Buttons Used in the Calibration Steps

The Meter buttons behave as follows when performing the Calibration Adjustment Procedure. This may be of help determining why a calibration step is not accepted and for determining the input value without referring to Table 16.

Press and hold to show the measured value. The measurement value is not

Acalibrated so it may not match the input value. This is normal. B Press and hold to display the required input amplitude.

GPress and hold to display the frequency of the required input.

Press to store the calibration value and advance to the next step. This button

Dis also used to exit calibration mode after the calibration adjustment sequence is complete.

Calibration Adjustment Procedure

Use the following steps to adjust the Meter's calibration. If the Meter is turned off before completion of the adjustment procedure, the calibration constants are not changed.

1.While holding down B, turn the rotary knob from OFF to VAC. The Meter displays “ZCAL”.

2.Press Donce to see the calibration counter.

3.Press Dagain to start the password entry. The Meter displays "????".

4.Press 4 buttons to enter the password.

5.Press Dto go to the first calibration step. The Meter displays "C-01" if the password is correct. If the password is not correct, the Meter emits a double beep, displays "????" and the password must be entered again. Repeat step 4.

6.Using Table 16, apply the input value listed for each calibration adjustment step. For each step, position the rotary switch and apply the input to the terminals as indicated in the table.

7.After each input value is applied, press Dto accept the value and proceed to the next step (C-02 and so forth).

Notes

After pressing D, wait until the step number advances before changing the calibrator source or turning the Meter rotary knob.

If the Meter rotary knob is not in the correct position, or if the measured value is not within the anticipated range of the input value, the Meter emits a double beep and will not continue to the next step.

Some adjustment steps take longer to execute than others (10 to 15 seconds). For these steps, the Meter will beep when the step is complete. Not all steps have this feature.

8.After the final step, the display shows "End" to indicate that the calibration adjustment is complete. Press Dto go to meter mode.