200.0 C Well temperature in degrees Celsius
SAccess set-point memory
1. 200. Set-point memory 1, 200.0°C currently used
Tochange the set-point memory press “UP” or “DOWN”.
5. 900. New set-point memory 5, 900.0°C
Press “SET” to accept the newselection and access the set-point value.
SAccept selected set-point memory
8.2.2 Set-point Value
The set-point valuemay be adjusted after selecting the set-point memory an d
pressing “SET”.
0900.0 Set-point value in °C
The firstdigit flashes on and off. If the set-point value is correct, press “EXIT”
to display the temperature scale units.
If the firstdigit is correct press “SET”.
Toadjust the first digit, press “UP” or “DOWN”. Then, press “SET” to go to
thesecond digit. The second digit of the temperature should now be flashing.
Adjustthis digit by pressing “UP” or “DOWN”.
Press “SET” to accept the second digit and repeat until the last digit has been
adjusted.
0962.7 New set-point value
Press “SET” to accept the newset-point. If “EXIT” is pressed all changes made
to the set-pointare discarded.
SAccept new set-point value
8.2.3 Temperature Scale Units
TemperatureScale Units of the controller are set by the user to degrees Celsius
(°C) or Fahrenheit(°F). The units are used in displaying the well temperature,
set-point,and proportional band.
Press “SET” afteradjusting the set-point value to change display units.
Un= C
23
8 Controller Operation
Temperature Set-point