troller cannot respond verywell to changing conditions or noise in the system.
If the proportionalband is too narrow the temperature may swing back and
forth because the controller overreactsto temperature variations. For best con-
trol stability the proportional band must be set for the optimum width.
The proportionalband widthis set at the factory to about 30.0°C. The propor-
tional band width may be altered by the user if he desires to optimize the con-
trol characteristics for a particular application.
The proportionalband width is easily adjusted from the front panel. The width
may be set todiscrete values in degrees C or F depending on the selected units.
The proportional band adjustmentcan be accessed within the secondary menu.
Press “SET”and “EXIT” to enter the secondary menu and show the heater
power.Then press “SET” twice to access the proportional band.
S+E Access heater power in secondary menu
100.0 P Heater power in percent
SAccess set-point voltage
11.934 U Set-point voltage in millivolts
SAccess proportional band
ProP Flashes and then displays the setting
30.0 Proportional band
Tochange the proportional band, press “UP” or “DOWN.”
30.5 New proportional band setting
Tostore the new setting press “SET”. Press “EXIT” to continue without storing
the newvalue.
SAccept the new proportional band setting
8.9 Controller Configuration
The controllerhas a number of configuration and operating options and calibra-
tion parameters which are programmable via the front panel. These are ac-
cessed from thesecondary menu after the proportional band function by
pressing “SET”. Pressing “SET” again entersthe first of three groups of config-
uration parameters—operating parameters, serial interfaceparameters and cali-
9150 Portable Furnace
User’s Guide
30