1200°C takes10 minutes to be within 0.5°C of its settled point and takes 15
minutes to achievemaximum stab ility.
Speeding up the calibration process can beaccomplished b y knowinghow soon
to make the measurement. It is recommended that typical measurements be
made at the desired temperatures with the desired test probes to establish these
times.
9150 Portable Furnace
User’s Guide
44