it can be adjusted.
00200 Program point 4 soak time set for 200 minutes
Press “SET” to savethe new soak-time value or “EXIT” to discard changes
SAccept the program point soak time
The nextvalue to edit is the program scan rate. This value is ignored if scan is
not enabled for the unit (See Section 8.3.1).
Sr 4 Program point 4 scan rate
Press “SET” to edit the program scan rate.
SEdit the program point scan rate
10.0 Current program point 4 scan rate
Use “UP” and “DOWN”to adjust the program scan rate.
11.3 New program point 4 scan rate
Press “SET” to savethe new scan rate value or “EXIT” to discard changes.
SAccept the program point scan rate
After “SET”is pressed the controller advances to the next program point or, if
thereare no more program pointsto edit, exits to the Program Function Menu.
Repeatthe above steps to edit any program point.
8.4.4 Program Function Mode
The nextparameter is the program function or cycle mode. There are four pos-
sible modes which determine whether the program scans up (from set-point 1
to n) only or both up and down(from set-point n to 1), and also whether the
program stops afterone cycle or repeats the cycle indefinitely. The table below
showsthe action of each of the four program mode settings.
Function Action
1 up-stop
2 up-down-stop
3 up-repeat
4 up-down-repeat
Pf=1
27
8 Controller Operation
Ramp and Soak Program