TROUBLESHOOTING
SYMPTOM | POSSIBLE CAUSES | REMEDIES |
|
|
|
Machine not operating | Thermal overload in motor | Allow the machine to cool: |
(continued) | has tripped | check the feed water supply |
|
| and/or amp draw to the |
|
| motor. |
|
|
|
| No power to machine | Check the fuses or circuit |
|
| breakers, measure the |
|
| voltage. |
|
|
|
| Motor and/or pump not oper- | See Tonkaflo pump instruc- |
| ating properly | tions. Contact GE |
|
| Osmonics for possible repair |
|
| or replacement. |
|
|
|
| Alarm condition has turned | Check for minimum inlet |
| OFF machine | pressure and push alarm |
|
| reset switch. |
|
|
|
Electrical machine shut | Alarm condition has turned | Restart the machine by |
down | OFF machine | pushing the alarm bypass. |
|
| Check for possible alarm |
|
| conditions: inlet pressure or |
|
| motor starter overload. |
|
|
|
| External control or permeate | Check for proper orientation |
| purge control contacts have | of control contacts. Restart |
| been shut machine down | machine manually or wait |
|
| for automatic |
|
|
|
30