| Table |
|
|
(6) a1 LED | This LED indicates that the test set is internally switched to the S11 or S21 mode and source power |
| is switched to Port 1. |
|
|
(7) a2 LED | This LED indicates that the test set is internally switched to the S22 or S12 mode and source power |
| is switched to Port 2. |
|
|
(8) Port 2 | This test port transmits RF energy from the source to the DUT and receives reflected or transmitted |
| RF energy from the DUT. |
|
|
|
|
(1) | Line Module | Houses the line cord connector, line fuse and line voltage selector. Pull out the right side |
|
| of the line module cover to replace or change the fuse or to change the voltage selection. |
|
| Remove the voltage selector drum to rotate it to a different voltage setting. Recommended |
|
| fuse values are printed on the rear panel. |
|
|
|
(2) | RF Input | A 2.4 mm connector that receives RF energy from the source. Connections made to this |
|
| input must be torqued no more than 90 |
|
|
|
(3) | J10 Test Set Interconnect | Used only in test sets with Option 001. It allows connecting another test set to the option |
|
| 001 test set. Up to four test sets can be serially connected to the analyzer. The HP 8510C |
|
| system automatically selects the IF output from the chosen test set for processing and |
|
| display. Refer to “Controlling Multiple Test Sets” in this chapter for more information. |
|
|
|
(4) | J11 Test Set Interconnect | Transmits the IF signal from the test set to the HP 85102 IF detector. It also transmits |
|
| control signals |
|
|
|
(5) | HP 8510 System Bus Address Switch | A |
|
| weight of each pole is indicated on the rear panel. So are the On and Off positions. The |
|
| default setting is decimal twenty |
|
|
|
HP 8517B