Measurement Calibration

 

After selecting the active test set, complete the system calibration procedure

 

as usual. When you select a different test set, make sure that you recall the

 

cal set that applies to that test set.

 

 

NOTE

Since the cal-set limited instrument state does not include the number of the

 

active test set, a cal set which does not apply to the current test set can be

 

turned on without displaying the HP 8510C caution messages. However, this

 

causes errors in the data displayed because incorrect error coefficients are

 

applied to the measured data.

 

For convenience, store a hardware-state file and an instrument-state file for

 

 

each combination of test set to cal set. You may also store your hardware

 

state file on a tape or disk for future use.

 

To change the configuration, recall the appropriate hardware state file. The

 

hardware state file:

sets the address of test set

issues the RF switch command, then

recalls the appropriate instrument state file which

recalls the cal set

HP 8517B S-Parameter Test Set Manual 3-8