Troubleshooting the Test Set

Troubleshooting Procedures

Testing Paths 5 and 6

1. Connect a cable between Port 1 and Port 2.

2. To observe the b1 power-level trace through path 5, press:

￿

PARAMETER [MENU]

￿

{USER 4 b1} then {REDEFINE PARAMETER}

￿

{DRIVE}

￿

DRIVE: {PORT 2}

￿

{PHASE LOCK}

￿

{a2}

￿

{REDEFINE DONE}

The b1 path signal is illustrated in Figure 5-12.Typically, each trace should resemble the examples in Figure 5-12,(±5 dB).

3.

Record your results in Figure 5-6.

4.

To observe the b2 power level trace through path 6, press:

 

￿

PARAMETER [MENU]

 

￿

{USER 2 b2} then {REDEFINE PARAMETER}

 

￿

{DRIVE}

 

￿

DRIVE: {PORT 1}

 

￿

{PHASELOCK}

 

￿

{a1}

 

￿

{REDEFINE DONE}

 

The b2 path signal is illustrated in Figure 5-12.Typically, each trace

 

should resemble the examples in Figure 5-12,(±5 dB).

 

5. Record your results in Figure 5-6.

 

 

NOTE

Be sure that you return the parameters to their original conditions for a1, b1,

 

a2, and b2. Use the redefine function to achieve this, or push the instrument

 

TEST switch to restore factory default conditions.

 

 

5-16HP 8517B S-Parameter Test Set Manual