Troubleshooting the Test Set
Troubleshooting Procedures
Testing Paths 5 and 6
1. Connect a cable between Port 1 and Port 2.
2. To observe the b1 | |
PARAMETER [MENU] | |
{USER 4 b1} then {REDEFINE PARAMETER} | |
{DRIVE} | |
DRIVE: {PORT 2} | |
{PHASE LOCK} | |
{a2} | |
{REDEFINE DONE} |
The b1 path signal is illustrated in Figure
3. | ||
4. | To observe the b2 power level trace through path 6, press: | |
| PARAMETER [MENU] | |
| {USER 2 b2} then {REDEFINE PARAMETER} | |
| {DRIVE} | |
| DRIVE: {PORT 1} | |
| {PHASELOCK} | |
| {a1} | |
| {REDEFINE DONE} |
| The b2 path signal is illustrated in Figure |
| |
| |
|
|
NOTE | Be sure that you return the parameters to their original conditions for a1, b1, |
| a2, and b2. Use the redefine function to achieve this, or push the instrument |
| TEST switch to restore factory default conditions. |
|
|