Troubleshooting the Test Set

Troubleshooting Procedures

Troubleshooting Procedures

The troubleshooting procedures in this chapter are listed below:

“A15 Primary Regulator Board Assembly” on page 5-7

“A27 Post-Regulator Board Assembly” on page 5-7

“Self-Test Indicators” on page 5-9

“HP 85102 IF Amplifier Test” on page 5-12

“Unratioed Power Test” on page 5-14

“HP 8517B, Option 007 Unratioed Power Test” on page 5-20

“Finding the Faulty Assembly” on page 5-23

A15 Primary Regulator Use a digital voltmeter to check the voltages and an oscilloscope to check for Board Assembly a maximum ripple level as listed in Table 5-2.

Table 5-2 Power Supply Voltages to A15

Nominal Voltage

Test Point

Voltage Range

Maximum Ripple

 

 

 

Peak-to-Peak

 

 

 

 

+5.05 Vdc

A15TP1

+4.75 to +5.25

2 mv

 

 

 

 

–5.20 Vdc

A15TP2

–4.90 to –5.50

2 mv

 

 

 

 

+14.85 Vdc

A15TP3

+14.10 to +15.60

2 mv

 

 

 

 

–14.85 Vdc

A15TP6

–14.10 to –15.60

2 mv

 

 

 

 

A27 Post-Regulator Board

(Option 007, Only) Use a digital voltmeter to check the voltages and an

Assembly

oscilloscope to check for a maximum ripple level as listed in Table 5-3.

Table 5-3 Power Supply Voltages to A27

 

 

 

 

 

 

 

 

Nominal Voltage

Test Point

Voltage Range

Maximum Ripple

 

 

 

 

Peak-to-Peak

 

 

 

 

 

 

+5.0 Vdc

A27TP1

+4.75 to +5.25

2 mv

 

 

 

 

 

 

–3.50 Vdc

A27TP2

–3.26 to –3.75

2 mv

 

 

 

 

 

HP 8517B S-Parameter Test Set Manual 5-7