Troubleshooting the Test Set

Troubleshooting Procedures

Figure 5-3 Power Supply Fuses and Test Points

HP-IB Address Switch The HP-IB address switch is located on the rear-panel of the instrument. It is easy to access, but unnecessary to change unless the error message SYSTEM BUS ADDRESS ERROR appears on the system display.

To set the HP-IB address switch, position the switches as illustrated in Figure 5-4(the LSB is on the left-hand side; the MSB is on the right-hand side). The default setting is decimal twenty, or binary 10100 as read from right-to-left (on, off, on, off, off). In the drawing, the dark-side is selected.