A6A4 Second Converter Replacement

Note

For location of hardware and cables referred to in this procedure, see Figure 4.

Removal

1.Remove ac line cords and RF Section bottom cover.

2.Disconnect connectors 0, @ and 0, single-pin connector 0, and double-pin connector @ from Second Converter.

3.Disconnect the remaining cables from Second Converter

1 (brown) cable from .321 OUT (7J

4 (yellow) cable from TUNE @

4. Remove two screws @ to remove Second Converter from A6 RF Module.

Installation

5.Position new Second Converter on mounting bracket. Secure with two screws 0.

6.Reconnect connector (ij, connector 0, single-pin connector 0, and double-pin connector @ to Second Converter.

7.Reconnect the following cables to Second Converter.

1 (brown) cable to .321 OUT 0

1 (brown) cable to .321 IN @

4 (yellow) cable to TUNE @

8. Replace bottom cover and ac line cords.