( 2 P L A C E S )
Figure 5. A6A6 First Converter Replacement
A6A15 3.6 GHz BPF, A6A16 LPF, and A6A18 LPF ReplacementNote | For location of hardware and cables referred to in this procedure, see Figure 6. |
3
Removal
1.Remove ac line cords, separate instrument sections, and remove top and bottom covers from RF Section.
2.Remove five cables Q) from A6 PC boards; three from A6A9 and two from A6A12.
3.Remove RF Module PC board cover by removing six screws 0.
4.Remove A6A9 assembly and A6A10, A6A11, and A6A12 PC boards.
5.Disconnect 4 (yellow) cable and 5 (green) cable from 2nd converter to release A6A9 assembly.
6.Disconnect coaxial cable @ between All YTO Loop and A6A5 ACLU at the ACLU.
7.Disconnect cable between A6A8 YTX and A6A5 ACLU at the YTX.
8.Disconnect cable between
9.Disconnect 1 (brown) cable from A6A4 2nd Converter .321 OUT.
10.Disconnect cable between A6A14 Limiter and A6Al RF Switch at the RF Switch.
11.Remove three screws @ used to attach ACLU cover plate one at a time, removing cable clamp and reinstalling the screws.
12.Disconnect wire harness connectors from ACLU and 2nd Converter to release wire harness from the “totem pole” assembly.
10 A6