7CleaningandMaintenance
7.1Cleaning
CAUTION!
Usecautiontopreventanyfluidfromenteringthe
CheckO2Plus.
–DONOTsoakorimmersetheCheckO2Plusinfluid.
–DONOTautoclaveorexposetheCheckO2Plusto
ethyleneoxidesterilization.
WipedowntheexteriorsurfacesoftheCheckO2Pluswithamoist
clothandmildhandordishsoap(pH6–8).
7.2Maintenance
•ReplacethebatterieswithhighqualityAAAlkalineorLithium
batteries.DONOTuserechargeablebatteries.
•Whennotinuseforperiodsgreaterthan30days,removethe
batteriestoprotecttheunitfrompotentialbatteryleakage.
•StoretheCheckO2Plusbetween–15°Cand60°C(5°F—
140°F)
141176119-A