Invacare FDX-CG MPJ+, PSR+, PSF+ Joysticks or Displays, Symptom Probable Cause Solutions

Models: FDX FDX-MCG FDX-CG

1 100
Download 100 pages 54.73 Kb
Page 94
Image 94

9 TROUBLESHOOTING

NUMBER OF

ERROR CODE

POSSIBLE SOLUTION

FLASHES

DESCRIPTION

 

 

 

 

7

Remote Fault

Check to make sure joystick is connected properly. Turn Joystick off then on.

 

 

Contact Invacare/Dealer for service.

 

 

 

8

Controller Fault

Contact Invacare/Dealer for service.

 

 

 

9

Communications Fault

Check joystick cable connections. Check joystick cable and connectors for damage.

 

 

Contact Invacare/Dealer for service.

 

 

 

10

General Fault

Contact Invacare/Dealer for service.

11Incompatible or Wrong type of remote connected. Contact Invacare/Dealer for service. incorrect Remote

MPJ™+, PSR+, PSF+ Joysticks or Displays

SYMPTOM

PROBABLE CAUSE

SOLUTIONS

 

 

 

LEFT BRAKE FAULT or RIGHT BRAKE

Motor lock levers disengaged (Error

Engage motor lock levers. Refer to Disengaging/Engaging

FAULT displays and wheelchair does not

code E9 or E10).

Motor Lock Levers on page 57.

drive.

 

 

 

 

 

CHARGER PLUGGED IN displays.

Battery charger connected (Error

Unplug battery charger from the wheelchair. Refer to

 

code E28).

Charging Batteries on page 53.

 

 

 

BATTERY FAULT displays and the wheelchair

Batteries need to be charged (Error

Charge batteries. Refer to Charging Batteries on page 53. If

does not drive.

code E14).

batteries fail to charge properly, check battery charger or

 

 

replace batteries. Refer to Replacing the Batteries and/or

 

 

Battery Cables on page 83.

 

 

 

JOYSTICK TIMEOUT displays and the

Joystick or input device is

Turn off power, reconnect the joystick of input device and

wheelchair does not drive.

disconnected (Error code 32).

turn power on.

 

 

 

Invacare® FDX®

94

Part No 1163181

Page 94
Image 94
Invacare FDX-CG, FDX-MCG user manual MPJ+, PSR+, PSF+ Joysticks or Displays, Symptom Probable Cause Solutions