Invacare FDX-CG Removing/Installing the Front Battery, Removing/Installing the Batteries

Models: FDX FDX-MCG FDX-CG

1 100
Download 100 pages 54.73 Kb
Page 79
Image 79

Removing/Installing the Front Battery

.

For this procedure, refer to FIGURE 1.

 

Reverse this procedure to install the front battery.

1. Place the wheelchair in a well ventilated area where work can be performed without risking damage to carpeting or floor covering.

2.If necessary, elevate the seat to the up most position.

3.Verify the joystick On/Off switch is in the Off position and disconnect joystick cable (not shown). Refer to Disconnecting/ Connecting the Joysticks on page 89.

4.Remove the two thumb screws securing the front shroud/battery retention bracket to the wheelchair base frame.

5.Remove the front shroud/battery retention bracket from the wheelchair base frame.

6.Disconnect the front battery connector from the front connector of the battery cable.

7.Remove the front battery from the wheelchair base frame.

Front Battery

Connector

8 SETUP/MAINTENANCE

Wheelchair Base Frame

Front Connector of Battery of

Battery Cable Located Here

Front Battery

 

Front Shroud/Battery

Mounting

Retention Bracket

Screws

FIGURE 1 Removing/Installing the Batteries

Part No 1163181

79

Invacare® FDX®

Page 79
Image 79
Invacare FDX-CG, FDX-MCG user manual Removing/Installing the Front Battery, Removing/Installing the Batteries