9 TROUBLESHOOTING

 

 

 

 

SYMPTOM

 

PROBABLE CAUSE

SOLUTIONS

 

 

 

 

JOYSTICK FAULT displays and the

 

The joystick or input device is

Replace joystick or input device.

wheelchair does not drive.

 

sending a value outside of the

 

 

 

reverse, forward, left or right limits

 

 

 

(Error codes E01, E02, E03 or E04).

 

 

 

 

 

NEUTRAL TESTING displays.

 

The joystick neutral test has failed

Release the joystick and try to get the joystick back into the

 

 

(Error code E18).

center-most position.

 

 

 

BAD JOYSTICK CAL VALUES displays and

The joystick calibration values are

Recalibrate the joystick (joystick throw procedure).

the wheelchair does not drive.

 

outside of the expected range

 

 

 

(Error code E19).

 

 

 

 

 

CTRL NOT CONNECTED

 

The MPJ or Display module is not

Check the connections between the joystick or display and

 

 

communicating with the control

the controller. Turn the power off and then back on. Replace

 

 

module (Error code E200).

the controller if necessary.

 

 

 

 

CTRL COM FAULT displays and the

 

The controller has determined a

Turn the wheelchair off and back on.

wheelchair drives slowly.

 

fault during a previous turn-off

 

 

 

process (Error code E41).

 

 

 

 

MISSING CONFIGURATION displays at

One or more devices have been

Reconnect the device.

power up.

 

removed or disconnected from the

 

 

 

wheelchair.

 

 

 

 

 

 

 

The Proportional or Digital

This is normal behavior.

ATTENDANT ACTIVE and

displays.

Attendant control is active and can

 

be used to drive the chair.

 

 

 

 

 

 

 

Batteries draw excessive current when

Battery failure.

Have batteries checked for shorted cell. Replace if necessary.

charging.

 

 

 

 

 

 

Contact Dealer/Invacare for service.

 

 

Electrical malfunction.

 

 

 

 

 

Part No 1163181

95

Invacare® FDX®

Page 95
Image 95
Invacare FDX-MCG, FDX-CG user manual Ctrl not Connected