8SETUP/MAINTENANCE

8.10Repositioning Joystick

For this procedure, refer to FIGURE 7.

This procedure must be performed by a qualified technician.

1.Turn the adjustment lock lever to release the joystick mounting tube from the mounting bracket.

2.Remove the joystick from the wheelchair.

3.Remove the three hex screws that secure both halves of the mounting bracket to the arm tube.

4.Reposition mounting bracket on opposite arm tube ensuring the threaded plate of the mounting bracket is on the inside of the arm tube as shown in FIGURE 7.

5.Using the three hex mounting screws and washers, secure both halves of the mounting bracket to the arm tube.

6.Slide the joystick mounting tube through the mounting bracket to the desired position.

7.Turn the adjustment lock lever to secure the joystick mounting tube into the mounting bracket.

 

Mounting Bracket

Mounting

Hex Mounting

Bracket

(Threaded

 

Screws

Plate)

 

 

Arm Tube

Washers

Adjustment Lock

Joystick

Lever

Mounting

 

Tube

Joystick not shown.

FIGURE 7 Repositioning Joystick