8 SETUP/MAINTENANCE

Check center mount front riggings for loose fasteners. Replace /tighten if necessary.

Check that cables are routed and secured properly to ensure that cables DO NOT become entangled and damaged during normal operation of seating system.

Ensure proper operation of powered functions (drive, seating, legrests, ect...).

Inspect motor brushes and gearbox coupling.

Inspect electrical components for signs of corrosion. Replace if corroded or damaged.

Inspect battery terminals for loose cable connection. Tighten if necessary.

Inspect all fasteners.

Inspect TRBKTS fasteners and hardware.

Inspect foam handgrips for damage. If damaged, have them replaced by a qualified technician.

Ensure swingarm stops are in place and not deteriorated or damaged. Replace if necessary.

Inspect/Adjust Every 18 Months

Replace motor brushes and gearbox coupling.

Part No 1163181